The Results of Long-term Follow-up of Total Hip Arthroplasty Using Hydroxyapatite-coated Cups.

Han Chang-Dong, Shin Keun-Young, Lee Hyun-Hee, Park Kwan-Kyu, Yang Ick-Hwan, Lee Woo-Suk

СтатьяПеревод в очереди

Аннотация

PURPOSE: The aim of this study was to report the long-term outcome and the failure mechanism of cementless total hip arthroplasty (THA) using hydroxyapatite (HA)-coated acetabular cup.

MATERIALS AND

METHODS

From January 1992 to May 1994, a total of 123 consecutive cementless primary THAs were performed using a HA-coated acetabular cup with metal-on-polyethylene articulation. We retrospectively evaluated 66 hips available for follow-up at a mean 18.3 years (range, 10.4-23.6 years). The survival analysis was performed by the Kaplan-Meier method. We defined end point as any failure that required a reoperation of acetabular component.

RESULTS

Thirty-nine of 66 hips (59.1%) were defined as a failure for progressive acetabular osteolysis or aseptic loosening of the cup. Acetabular osteolysis was observed in 47 hips (71.2%) and 33 hips (50.0%) were revised because of cup loosening. The Kaplan-Meier method showed the survival rate of the acetabular cup to be 46.3% at 15 years and 34.8% at 20 years for any failure that required a reoperation of acetabular component.

CONCLUSION

The long-term survival rate of THA using HA-coated acetabular cup was unsatisfactory, and it was attributed to vulnerable property of HA coating and progressive osteolysis.

Для цитирования: Han Chang-Dong, Shin Keun-Young, Lee Hyun-Hee, Park Kwan-Kyu, Yang Ick-Hwan, Lee Woo-Suk. The Results of Long-term Follow-up of Total Hip Arthroplasty Using Hydroxyapatite-coated Cups.. Hip & pelvis. 2015. DOI: 10.5371/hp.2015.27.4.209.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.