Lead performance of stylet-driven leads in left bundle branch area pacing: Results from a large single-center cohort and insights from in vitro bench testing.

Özpak Emine, Van Heuverswyn Frederic, Timmermans Frank, De Pooter Jan

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Left bundle branch area pacing (LBBAP) requires deep septal lead deployment for left-sided conduction stimulation. Advancing leads toward deep septal positions might add mechanical stress on these leads. Concerns about lead performance and reliability remain an unanswered question.

OBJECTIVE

The purpose of this study was to analyze lead performance and integrity of stylet-driven pacing leads (SDLs) for LBBAP.

METHODS

This study assessed lead fracture rates of SDL in a large single-center cohort of adult LBBAP patients. Fluoroscopic analysis of lead bending angulations at the septal insertion point and in vitro bench testing of lead preconditioning were performed to simulate clinical use conditions. Lead performance was compared between LBBAP and conventional right ventricular apical pacing (RVp) sites.

RESULTS

The study included 325 LBBAP patients (66% male; age 71±15 years). During median follow-up of 18 months, 2 patients (0.6%) experienced conductor fracture between tip housing and ring electrode, whereas no such fractures occurred with RVp patients (n = 149; P = .22). X-ray analysis revealed that high lead bending angulations occurred in 1.3% of the patients. Accelerated bench testing of excessive preconditioned leads showcased a higher probability of early conductor fracture compared to standard preconditioned leads.

CONCLUSION

The incidence of early conductor failure in LBBAP seems higher than with conventional RVp sites. The most vulnerable lead part seems to be the interelectrode space between the tip housing and ring electrode. Excessive angulation and preconditioning might contribute to early fatigue fracture.

Для цитирования: Özpak Emine, Van Heuverswyn Frederic, Timmermans Frank, De Pooter Jan. Lead performance of stylet-driven leads in left bundle branch area pacing: Results from a large single-center cohort and insights from in vitro bench testing.. Heart rhythm. 2024. DOI: 10.1016/j.hrthm.2024.01.049.