Comparison of pH, lactate, and glucose analysis of equine synovial fluid using a portable clinical analyzer with a bench-top blood gas analyzer.

Dechant Julie E, Symm William A, Nieto Jorge E

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To compare agreement between a portable clinical analyzer and laboratory-based bench-top analyzer for analysis of pH, lactate, and glucose concentrations in synovial fluid.

STUDY

DESIGN

Prospective experimental study.

ANIMALS: Clinically normal horses (n=8); 6 horses euthanatized for reasons unrelated to the study; 11 horses that had synoviocentesis for reasons other than sepsis; 7 horses that had synoviocentesis for evaluation of sepsis; and 2 horses without recorded clinical data. Median age of horses was 8 years (range, 1 day to 24 years).

METHODS

Supernatant from each synovial fluid sample was analyzed for pH, lactate, and glucose concentrations using an ABL 705 laboratory-based bench-top analyzer and i-STAT portable clinical analyzer. Bland-Altman plots were constructed and concordance analysis performed to determine bias and agreement between the 2 analyzers.

RESULTS

There was acceptable agreement between analyzers for lactate and glucose concentrations, with biases of 0.198 mmol/L and 9 mg/dL and concordance correlation coefficients of 0.97 and 0.96 for lactate and glucose, respectively. The agreement between analyzers for pH was not acceptable, with a bias of -0.057 and concordance correlation coefficient of 0.89.

CONCLUSIONS

This study found that the portable clinical analyzer performed similarly to the bench-top blood gas analyzer for evaluation of lactate and glucose concentrations, but not pH, in synovial fluid.

Для цитирования: Dechant Julie E, Symm William A, Nieto Jorge E. Comparison of pH, lactate, and glucose analysis of equine synovial fluid using a portable clinical analyzer with a bench-top blood gas analyzer.. Veterinary surgery : VS. 2011. DOI: 10.1111/j.1532-950x.2011.00854.x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.