Poor performance of Enduron polyethylene liner in total hip arthroplasty: a minimum ten-year follow up and ultra-morphological analysis of wear particles.

Yan Yufei, Chen Hao, Feng Jianmin, Chen Kaizhe, Zhou Kaidi, Hong Weixiang, Wang Yi, Liu Zhihong, Zhang Jiong, Yang Qingming, Guo Lei, He Chuan

Clinical StudyСтатьяПеревод в очереди

Аннотация

PURPOSE: The aim of the present study was to investigate the long-term outcome and the wear characteristics of two distinct types of ultra-high molecular weight polyethylene (UHMWPE) liners in total hip arthroplasty (THA).

METHODS

We conducted a retrospective clinical study on patients which were treated with total hip arthroplasty using either Enduron polyethylene (Enduron PE) or Trilogy polyethylene (Trilogy PE) liners based on a minimum of ten year follow up data. Morphological analyses of wear particles from tissue samples, which were harvested during revision surgeries, were also performed.

RESULTS

A total of 79 THAs in the Enduron group and 55 THAs in the Trilogy group were available for analysis. Kaplan-Meier survival with revision for wear-related complications as the endpoint of the Enduron PE liners was lower than that of Trilogy PE liners at ten years (93.5 % and 100 %, P = 0.03). The Enduron group had higher mean linear wear rate than that of the Trilogy group (0.20 ± 0.09 and 0.09 ± 0.03 mm/year, P < 0.01). The incidence of osteolysis for the Enduron group was higher than that of the Trilogy group (33.3 % and 12 %, P = 0.04). Under transmission electron microscopy, the Enduron group had more than 82 % of the particles less than 1.0 μm in size and more than 57 % of the particles less than 0.5 μm.

CONCLUSION

The long-term performance of Enduron liners was worse than that of Trilogy liners. Further clinical follow-up may be necessary in patients with Enduron PE liners in order to avoid catastrophic complications.

Для цитирования: Yan Yufei, Chen Hao, Feng Jianmin, Chen Kaizhe, Zhou Kaidi, Hong Weixiang, Wang Yi, Liu Zhihong, Zhang Jiong, Yang Qingming, Guo Lei, He Chuan. Poor performance of Enduron polyethylene liner in total hip arthroplasty: a minimum ten-year follow up and ultra-morphological analysis of wear particles.. International orthopaedics. 2016. DOI: 10.1007/s00264-016-3217-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.