[A comparative study of Taylor spatial frame and unilateral external fixator in treatment of tibiofibular open fractures].

Ma He, Yao Hui, Zhang Tao, Wan Chunyou

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

OBJECTIVE

To compare the effectiveness of Taylor spatial frame (TSF) and unilateral external fixator in the treatment of tibiofibular open fractures.

METHODS

Between January 2016 and July 2018, 74 patients with tibiofibular open fracture who met the selection criteria were divided into TSF group (43 cases, fixed with TSF) and unilateral group (31 cases, fixed with unilateral external fixator) according to the principle of entering the group every other day. There was no significant difference in gender, age, affected side, cause of injury, type of fracture between the two groups ( P>0.05). The operation time, fracture healing time, removal time of external fixator, and complications were recorded and compared between the two groups. The limb function was evaluated according to Johner-Wruhs criteria for evaluating the final effectiveness of tibial shaft fracture treatment. The recovery of lower limb force line was ecaluated by LUO Congfeng et al. criteria.

RESULTS

All patients were followed up 8-22 months, with a median of 12 months. All fractures healed, and no complication such as delayed union, nonunion, or osteomyelitis occurred. The operation time, fracture healing time, and removal time of external fixator in TSF group were significantly shorter than those in unilateral group ( P<0.05). At 3 months after the removal of the external fixator, the limb function was evaluated according to the Johner-Wruhs standard. In TSF group, 41 cases were excellent, 1 case was good, and 1 case was fair, and the excellent and good rate was 97.67%; in unilateral group, 30 cases were excellent and 1 case was fair, and the excellent and good rate was 96.77%; there was no significant difference between the two groups ( P=0.666). At 4 months after operation, the recovery of lower limb force line was ecaluated by LUO Congfeng et al. criterion. In TSF group, 41 cases were excellent, 2 cases were good, and 1 case was fair, and the excellent and good rate was 97.67%; in unilateral group, 29 cases were excellent, 1 case was good, 1 case was fair, and the excellent and good rate was 96.77%; there was no significant difference between the two groups ( P=0.666).

CONCLUSION

For tibiofibular open fracture, on the premise of fracture healing, TSF technology is superior to unilateral external fixation in terms of shortening operation time, fracture healing time, and removal time of external fixator.

Для цитирования: Ma He, Yao Hui, Zhang Tao, Wan Chunyou. [A comparative study of Taylor spatial frame and unilateral external fixator in treatment of tibiofibular open fractures].. Zhongguo xiu fu chong jian wai ke za zhi = Zhongguo xiufu chongjian waike zazhi = Chinese journal of reparative and reconstructive surgery. 2020. DOI: 10.7507/1002-1892.201909120.