MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/pm/pny245

Медицинская библиотека · Pain medicine (Malden, Mass.) · 2019

Explantation of Percutaneous Spinal Cord Stimulator Devices: A Retrospective Descriptive Analysis of a Single-Center 15-Year Experience.

Simopoulos Thomas, Aner Moris, Sharma Sanjiv, Ghosh Priyanka, Gill Jatinder S

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

The aims of this study were to identify the reasons for explanation of spinal cord stimulator (SCS) devices and to further quantify the proportions and time lines of these causes of explantation in order to determine improvement opportunities for the development of best practices.

DESIGN

Retrospective, single academic center.

MATERIALS AND

METHODS

Patients who were implanted with percutaneous SCS devices from 2002 to 2015 and with follow-up available until the end of September 2017 were included in this retrospective chart analysis.

RESULTS

Of the 356 patients trialed, 252 underwent implantation of an SCS device with a permanent to trial ratio of 71%. Of the patients who had a permanent implant, 50% had failed back surgery syndrome, 25% had complex regional pain syndrome, and 25% had other diagnoses. At the end of the study period, the explantation rate was 30%. The causes for explantation included biological complications (26.6%), paresthesia limitations or side effects (26.6%), hardware complications (13.3%), ineffective pain control (28%), and no further need for stimulation therapy (5.3%).

CONCLUSIONS

Device removal is not uncommon, and opportunities to enhance the long-term success of SCS devices do exist. These include modification of trialing methods, achieving better paresthesia overlay, using magnetic resonance imaging (MRI) conditional equipment, development of robust technologies and hardware to reduce equipment malfunction, and improving efficacy with new innovative wave forms.

Для цитирования: Simopoulos Thomas, Aner Moris, Sharma Sanjiv, Ghosh Priyanka, Gill Jatinder S. Explantation of Percutaneous Spinal Cord Stimulator Devices: A Retrospective Descriptive Analysis of a Single-Center 15-Year Experience.. Pain medicine (Malden, Mass.). 2019. DOI: 10.1093/pm/pny245.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.