MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1055/s-0043-1771198

Медицинская библиотека · Veterinary and comparative orthopaedics and traumatology : V.C.O.T · 2023

Effect of Plate-Bone Distance and Working Length on 2.0-mm Locking Construct Stiffness and Plate Strain in a Diaphyseal Fracture Gap Model: A Biomechanical Study.

Evans Alecia, Glyde Mark, Day Rob, Hosgood Giselle

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

The aim of this study was to determine the effect of plate-bone distance (PBD) and working length on 2.0-mm locking compression plate (LCP) stiffness and strain in four-point bending and torsion in a diaphyseal fracture gap model.

STUDY

DESIGN

A total of 54 LCP with three screws per fragment were assigned to one of nine combinations of working length (WL; short, medium, and long), and PBD (1, 1.5, and 3 mm) for a sample size of six per construct configuration. Stiffness was measured under quasistatic, nondestructive four-point compression bending and torsion. Plate surface strain was recorded using three-dimensional (3D) digital image correlation during four-point compression bending.

RESULTS

WL had a significant effect on overall construct stiffness in both compression bending and in torsion, with shorter WL constructs having higher stiffness (p < 0.0001). PBD had no effect on construct stiffness in compression bending; however, a significant reduction in stiffness was noted in torsion (p = 0.047) as PBD incrementally increased. WL had a significant effect on plate strain in compression bending, with shorter WL constructs having lower plate strain (p < 0.0001). PBD had no effect on plate strain in compression bending except for lower plate strain recorded in long WL constructs with 1-mm PBD, compared with 1.5- and 3-mm PBD constructs (p < 0.0001).

CONCLUSION

Longer WL constructs, regardless of PBD, had lower stiffness in compression bending, while in torsion, some modulation of this effect was noted with incremental decreases in PBD. Longer WL resulted in high plate strain, regardless of PBD.

Для цитирования: Evans Alecia, Glyde Mark, Day Rob, Hosgood Giselle. Effect of Plate-Bone Distance and Working Length on 2.0-mm Locking Construct Stiffness and Plate Strain in a Diaphyseal Fracture Gap Model: A Biomechanical Study.. Veterinary and comparative orthopaedics and traumatology : V.C.O.T. 2023. DOI: 10.1055/s-0043-1771198.