Mid term results of total hip arthroplasty using polyethylene-ceramic composite (Sandwich) liner.

Wang Tao, Sun Jun-Ying, Zha Guo-Chun, Dong Sheng-Jie, Zhao Xi-Jiang

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Ceramic-on-ceramic (COC) couplings are an attractive alternative bearing surfaces that have been reported to eliminate or reduce problems related to polyethylene wear debris. However, the material in total hip arthroplasty (THA) remains one of the major concern regarding the risk of fracture. The present study aims at reporting the fracture rate of bearings in a series of COC THAs with the use of a sandwich liner and attempt to detect the relative risk factors, the possible cause and assess the clinical results.

MATERIALS AND

METHODS

We retrospectively evaluated 153 patients (163 hips) using the sandwich liner COC THA between 2001 and 2009. Patient assessment was based on demographic factors, including age, weight, gender and body-mass index (BMI). All patients were evaluated clinically and radiographically or using computed tomography viz-a-viz dislocation, osteolysis, periprosthetic fracture, infection, loosening and implant fracture.

RESULTS

Three ceramic sandwich liners fracture (1.84%) were observed at an average of 7.3 years' followup. The factors which were found to be non-significant to the ceramic liner fracture, included age (P = 0.205), weight (P = 0.241), gender (P = 0.553), BMI (P = 0.736), inclination (P = 0.199) and anteversion (P = 0.223). The overall survival was 91.4% at 12-year with revision as the endpoint. Other complications included osteolysis in 4 (2.45%), dislocation in one and periprosthetic fracture in one. In no hip aseptic loosening of the implants was seen.

CONCLUSIONS

Our experience with the ceramic-polyethylene sandwich liner acetabular component has been disappointing because of the high rate of fracture and osteolyis. We have discontinued the use of this device and recommend the same.

Для цитирования: Wang Tao, Sun Jun-Ying, Zha Guo-Chun, Dong Sheng-Jie, Zhao Xi-Jiang. Mid term results of total hip arthroplasty using polyethylene-ceramic composite (Sandwich) liner.. Indian journal of orthopaedics. 2016. DOI: 10.4103/0019-5413.173512.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.