DOI 10.1097/rli.0000000000001085
Postoperative Extremity Tomosynthesis-A Superimposition-Free Alternative to Standard Radiography?
Аннотация
RATIONALE AND
OBJECTIVES
This study investigates the performance of tomosynthesis in the presence of osteosynthetic implants, aiming to overcome superimposition-induced limitations in conventional radiograms.
MATERIALS AND
METHODS
After surgical fracture induction and subsequent osteosynthesis, 8 cadaveric fracture models (wrist, metacarpus, ankle, metatarsus) were scanned with the prototypical tomosynthesis mode of a multiuse x-ray system. Tomosynthesis protocols at 60, 80, and 116 kV (sweep angle 10°, 13 FPS) were compared with standard radiograms. Five radiologists independently rated diagnostic assessability based on an equidistant 7-point scale focusing on fracture delineation, intra-articular screw placement, and implant positioning. The intraclass correlation coefficient (ICC) was calculated to analyze interrater agreement.
RESULTS
Radiation dose in radiography was 0.48 ± 0.26 dGy·cm 2 versus 0.12 ± 0.01, 0.36 ± 0.02, and 1.95 ± 0.11 dGy·cm 2 for tomosynthesis scans at 60, 80, and 116 kV. Delineation of fracture lines was superior for 80/116 kV tomosynthesis compared with radiograms ( P ≤ 0.003). Assessability of intra-articular screw placement was deemed favorable for all tomosynthesis protocols ( P ≤ 0.004), whereas superiority for evaluation of implant positioning could not be ascertained (all P 's ≥ 0.599). Diagnostic confidence was higher for 80/116 kV tomosynthesis versus radiograms and 60 kV tomosynthesis ( P ≤ 0.002). Interrater agreement was good for fracture delineation (ICC, 0.803; 95% confidence interval [CI], 0.598-0.904), intra-articular screw placement (ICC, 0.802; 95% CI, 0.599-0.903), implant positioning (ICC, 0.855; 95% CI, 0.729-0.926), and diagnostic confidence (ICC, 0.842; 95% CI, 0.556-0.934).
CONCLUSIONS
In the postoperative workup of extremity fractures, tomosynthesis allows for superior assessment of fracture lines and intra-articular screw positioning with greater diagnostic confidence at radiation doses comparable to conventional radiograms.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.