DOI 10.1097/01.brs.0000239132.16484.be

Медицинская библиотека · Spine · 2006

Wear and corrosion in retrieved thoracolumbar posterior internal fixation.

Villarraga Marta L, Cripton Peter A, Teti Stephanie D, Steffey Duane L, Krisnamuthy Saki, Albert Todd, Hilibrand Alan, Vaccaro Alexander

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

STUDY

DESIGN

Posterior thoracolumbar spine implants retrieved as part of routine clinical practice over a 2-year period were analyzed to identify wear and corrosion.

OBJECTIVE

Engineering analyses of retrieved posterior instrumentation for indications of performance and failure and correlation of this information with clinical factors.

SUMMARY OF BACKGROUND DATA: Recent studies have reported spinal instrumentation particulate wear debris and have noted the importance of design considerations at implant connector interfaces.

METHODS

A total of 57 implants were analyzed from patients (39 female, 18 male) whose average age at implantation was 43.9 years (range, 13.7-77.4 years). Time of implantation ranged from 2 months to 13.5 years. The top 3 implantation diagnoses were radiculopathy (33%), scoliosis (30%), and back pain (25%). Metallurgical analyses were performed to characterize the wear and/or corrosion, and fractures of the implants.

RESULTS

Wear was present in 75%, corrosion in 39%, and fractures in 7% of the retrieved implants. Wear and/or corrosion was more prevalent, with respect to the total number of implants retrieved, in implants that had been in service at least 1 year. There was no evidence of corrosion in any of the Ti implants, whereas corrosion was present (with wear) in 58% of the stainless steel (SS) implants. Wear and corrosion were more frequently observed in long rods than in short rods. Implantation times were longer for SS implants than for Ti implants.

CONCLUSIONS

Retrieved rods exhibited corrosion, wear, and fracture, with wear and corrosion mainly located at the interfaces with hooks, screws, or cross-connectors. The mechanisms causing this material loss in situ, as well as what local or systemic responses it may stimulate are of clinical significance and should be studied further.

Для цитирования: Villarraga Marta L, Cripton Peter A, Teti Stephanie D, Steffey Duane L, Krisnamuthy Saki, Albert Todd, Hilibrand Alan, Vaccaro Alexander. Wear and corrosion in retrieved thoracolumbar posterior internal fixation.. Spine. 2006. DOI: 10.1097/01.brs.0000239132.16484.be.

Предметные рубрики MeSH