Медицинская библиотека · The Journal of bone and joint surgery. American volume · 2003

Serum cobalt levels after metal-on-metal total hip arthroplasty.

Brodner Wolfram, Bitzan Peter, Meisinger Vanee, Kaider Alexandra, Gottsauner-Wolf Florian, Kotz Rainer

Клиническое исследованиеСтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Systemic cobalt dissemination from the Metasul Co-28Cr-6Mo-0.2C metal-on-metal total hip prosthesis has been demonstrated in the first year after implantation. The aim of this prospective study was to monitor the serum cobalt concentrations in patients during the first five years after total hip arthroplasty with a metal-on-metal articulation.

METHODS

A total hip arthroplasty was performed without cement in 100 consecutive patients who had either unilateral osteoarthritis or unilateral osteonecrosis. Fifty patients were randomized to be treated with a metal-on-metal articulation, and fifty patients, with a ceramic-on-polyethylene bearing. The femoral stem was made of a Ti-6Al-7Nb alloy, and the threaded acetabular cup was made of commercially pure titanium. Blood samples were taken before the operation and at multiple time-points for five years after the operation. Serum cobalt concentrations were measured with use of atomic absorption spectrometry.

RESULTS

In the metal-on-metal group, the median serum cobalt concentration was 1 micro g/L at one year after surgery and 0.7 micro g/L at five years. The median of the serum cobalt concentrations measured from three to twelve months did not differ from the median of subsequent measurements, with the numbers available. The median serum cobalt level in the control group of patients treated with the ceramic-on-polyethylene articulation was below the detection limit at all time-points.

CONCLUSIONS

Systemic cobalt release from Metasul metal-on-metal articulations was demonstrated throughout the five-year study period. The median serum cobalt concentrations were found to be slightly above the detection limit and remained in a constant range. The serum cobalt concentrations did not reflect a so-called run-in wear period of the metal-on-metal articulations.

Для цитирования: Brodner Wolfram, Bitzan Peter, Meisinger Vanee, Kaider Alexandra, Gottsauner-Wolf Florian, Kotz Rainer. Serum cobalt levels after metal-on-metal total hip arthroplasty.. The Journal of bone and joint surgery. American volume. 2003. DOI: 10.2106/00004623-200311000-00017.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.