Independent Evaluation of a Mechanical Hip Socket Navigation System in Total Hip Arthroplasty.

Jennings Jason M, Randell Timmothy R, Green Cynthia L, Zheng Guoyan, Wellman Samuel S

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Acetabular cup malpositioning during total hip arthroplasty may lead to impingement, instability, wear-induced osteolysis, and increased rates of revision surgery. The purpose of this study was to independently evaluate the accuracy of acetabular cup orientation using a novel mechanical navigation device.

METHODS

An adjustable, reusable mechanical navigation device was used in a consecutive series of patients. Angles (inclination and anteversion) were measured by 2 independent reviewers with a validated 2-dimensional/3-dimensional matching application using a preoperative computed tomographic scan and a postoperative plain film.

RESULTS

There were no outliers for inclination or anteversion for errors within 10° of the preoperative plan. There were 6 (12.8%) outliers for inclination and 11 (23.4%) for anteversion for errors within 5° of the preoperative plan. All patients simultaneously met both targets when an outlier was considered 10°, whereas 30 (63.8%) simultaneously met both targets when the criteria was tightened to 5°. The absolute errors for both inclination and anteversion were significantly less than both 5° and 10° (P<.001).

CONCLUSION

This device reliably navigates acetabular cup inclination and anteversion individualized for each patient to within 10° of a preoperative plan. Outliers increased when this criteria is tightened to 5° but still appears to be more accurate than conventional acetabular cup component placement. Further research is warranted to assess the clinical impact of reducing outliers using this device.

Для цитирования: Jennings Jason M, Randell Timmothy R, Green Cynthia L, Zheng Guoyan, Wellman Samuel S. Independent Evaluation of a Mechanical Hip Socket Navigation System in Total Hip Arthroplasty.. The Journal of arthroplasty. 2015. DOI: 10.1016/j.arth.2015.09.052.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.