DOI 10.1097/j.pbj.0000000000000208
Total knee arthroplasty coronal alignment and tibial base stress-a new numerical evaluation.
Аннотация
BACKGROUND
Total knee arthroplasty (TKA) is one of the most frequently performed orthopedic procedures. The correct positioning and alignment of the components significantly affects prosthesis survival. Considering the current controversy regarding the target of postoperative alignment of TKA, this study evaluated the tension at tibial component interface using two numerical methods.
METHODS
The stress of the prosthesis/bone interface of the proximal tibial component was evaluated using two numerical methods: the finite element method (FEM) and the new meshless method: natural neighbor radial point interpolation method (NNRPIM). The construction of the model was based on Zimmers NexGen LPS-Flex Mobile® prosthesis and simulated the forces by using a free-body diagram.
RESULTS
Tibiofemoral mechanical axis (TFMA) for which a higher number of nodes are under optimal mechanical tension is between 1° valgus 2° varus. For values outside the interval, there are regions under the tibial plate at risk of bone absorption. At the extremities of the tibial plate of the prosthesis, both medial and lateral, independent of the alignment, are under a low stress. In all nodes evaluated for all TFMA, the values of the effective stresses were higher in the NNRPIM when compared with the FEM.
CONCLUSION
Through this study, we can corroborate that the optimal postoperative alignment is within the values that are currently considered of 0 ± 3° varus. It was verified that the meshless methods obtain smoother and more conservative results, which may make them safer when transposed to the clinical practice.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.