MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/art.22612

Медицинская библиотека · Arthritis and rheumatism · 2007

Computed tomography measurement of tophus volume: comparison with physical measurement.

Dalbeth Nicola, Clark Barnaby, Gregory Kate, Gamble Gregory D, Doyle Anthony, McQueen Fiona M

Сравнительное исследованиеОценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

Computed tomography (CT) has high accuracy for tophus detection. This study assessed reliability of CT measurement of tophus volume and compared reproducibility of CT with physical measurement of tophus size.

METHODS

Forty-seven hand tophi were analyzed in 20 patients with gout. The longest tophus diameter was recorded by 2 independent observers. All patients proceeded to CT scanning of the hands on a Philips Brilliance scanner (0.8-mm slices). Two independent observers measured tophus volume using the Surface Shaded Display 3-dimensional function on the Philips CT workstation. Five patients underwent repeat physical and CT assessments within 1 week (18 observations). Inter- and intraobserver reproducibility were analyzed by limits of agreement and coefficients of variation.

RESULTS

Of the 47 lesions identified as tophi on physical examination, 42 (89%) were also identified on CT. The mean (95% confidence interval [95% CI]) difference between observers for physical measurement was 0.45 mm (-4.07, 4.96) and for CT was 65.2 mm(3) (-293.0, 423.3). The mean (95% CI) difference between visits for physical measurement was -0.72 mm (-5.47, 4.03) and for CT was -13.1 mm(3) (-112.5, 86.3). There was no difference between coefficients of variation for inter- and intraobserver reproducibility for the 2 measurement techniques. For tophi identified by physical and CT assessment, there was good correlation between measurements (r = 0.91, P < 0.0001).

CONCLUSION

CT assessment of tophus volume is reliable and reproducible. However, physical measurement correlates well with CT and has equivalent reproducibility. These data support the use of physical measurement as a simple and reliable method to assess tophus size.

Для цитирования: Dalbeth Nicola, Clark Barnaby, Gregory Kate, Gamble Gregory D, Doyle Anthony, McQueen Fiona M. Computed tomography measurement of tophus volume: comparison with physical measurement.. Arthritis and rheumatism. 2007. DOI: 10.1002/art.22612.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.