Performance of the Aktiia optical blood pressure measurement device in the elderly: a comparison with double blinded auscultation in different body positions.

Theiler Kenji, Sola Josep, Damianaki Aikaterini, Pfister Alexandre, Almeida Tiago P, Alexandre Jérémy, Vermare Pascale, Wuerzner Gregoire

Рандомизированное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Accurate blood pressure (BP) measurement is essential for the correct diagnosis and management of hypertension (HTN) especially in the elderly population. As with of all BP devices, the accuracy of cuffless devices must be verified. This study (NCT04027777) aimed to evaluate the performance of a wrist cuffless optical BP device in an elderly population cohort in different body positions with auscultation as the reference measurement.

DESIGN AND

METHODS

Patients aged 65-85 years with different BP categories but without diabetes were recruited. After an initial calibration based on auscultatory measurements, BP estimation from the Aktiia Bracelet (Aktiia SA, Switzerland) were compared to reference double-blinded auscultatory measurements in sitting, standing and lying positions on four separate visits distributed over one month. In the absence of a universal standard for cuffless BP device at the time of the study, modified ISO81060-2 criteria were used for performance analysis.

RESULTS

Thirty-five participants were included in the analysis fulfilling the inclusion requirements of ISO 81060-2. A total of 469 paired measurements were obtained with overall 83% acceptance rate. Differences (mean ± SD)   between Aktiia Bracelet and auscultation for systolic BP were -0.26 ± 9.96 mmHg for all body positions aggregated (sitting 1.23 ± 7.88 mmHg, standing -1.81 ± 11.11 mmHg, lying -1.8 ± 9.96 mmHg). Similarly, differences for diastolic BP were -0.75 ± 7.0 mmHg (0.2 ± 5.55 mmHg, -5.35 ± 7.75 mmHg and -0.94 ± 7.47 mmHg, respectively). Standard deviation of the averaged differences per subject for systolic/diastolic BP was 3.8/2.5 mmHg in sitting and 4.4/3.7 mmHg for all body positions aggregated.

CONCLUSIONS

Overall, this study demonstrates a similar performance of the Aktiia Bracelet compared to auscultation in an elderly population in body positions representative of daily activities. The use of more comfortable, non-invasive, and non-occlusive BP monitors during long periods may facilitate e-health and may contribute to better management of HTN, including diagnosis and treatment of HTN, in the elderly.

Для цитирования: Theiler Kenji, Sola Josep, Damianaki Aikaterini, Pfister Alexandre, Almeida Tiago P, Alexandre Jérémy, Vermare Pascale, Wuerzner Gregoire. Performance of the Aktiia optical blood pressure measurement device in the elderly: a comparison with double blinded auscultation in different body positions.. Blood pressure. 2023. DOI: 10.1080/08037051.2023.2281320.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.