DOI 10.1097/bot.0000000000002594
Effect of Implant Linkage on Axial and Rotational Stiffness of Nail-Plate Constructs for Comminuted Distal Femoral Fractures.
Аннотация
OBJECTIVES
To determine the effect of linking the intramedullary nail and the laterally applied locking plate to treat comminuted distal femur fractures and allow for immediate weight bearing.
METHODS
Comminuted extra-articular distal femur fractures were created in 16 synthetic osteoporotic femurs and split into 2 groups: linked versus unlinked. In the linked construct, in addition to standard plate fixation and proximal locking of the nail, 2 nonthreaded locking bolts (prototypes) were placed through both the plate and the nail. In the unlinked construct, the same number of screws affixed the plate to the bone but were directed around the nail, and separate distal interlocking screws were placed for nail fixation. Each specimen underwent sequential axial and torsional loading, and axial and torsional stiffness were calculated and compared.
RESULTS
On average, the unlinked constructs demonstrated a greater axial stiffness at all levels of axial loading, and linked constructs demonstrated greater average rotational stiffness. However, there were no statistically significant differences ( P > 0.189) between the linked and unlinked groups at any axial or torsional load.
CONCLUSION
In distal femur fractures with metaphyseal comminution, there were no significant differences in axial or torsional stiffness when linking the plate to the nail. Linking the construct seems to provide no significant mechanical advantage compared with the unlinked environment, but it may be a useful strategy to reduce nail "traffic" in the distal segment with no significant disadvantage.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.