Ultrasound velocity through the cortex of phalanges, radius, and tibia in normal and osteoporotic postmenopausal women using a new multisite quantitative ultrasound device.

Damilakis John, Papadokostakis George, Vrahoriti Hrisanthi, Tsagaraki Ioanna, Perisinakis Kostas, Hadjipavlou Alexander, Gourtsoyiannis Nicholas

СтатьяПеревод в очереди

Аннотация

RATIONALE AND

OBJECTIVES

To assess a new multisite quantitative ultrasound (QUS) device (Sunlight Omnisense 7000 S) suitable for the measurement of speed of sound (SOS) in the phalanges, radius, and tibia.

METHODS

The study group consisted of 270 healthy Caucasian postmenopausal patients (mean age: 60.0 +/- 7.6 years) and 53 Caucasian postmenopausal patients (mean age: 67.2 +/- 7.4 years) with osteoporotic fractures. Measurements of SOS and bone mineral density (BMD) were carried out in all subjects.

RESULTS

Intraobserver in vivo short-term precision was on average 0.76% for the radius, 0.47% for the tibia, and 1.54% for the phalanges. The interobserver precision ranged from 0.77% to 2.39%. Measurements of SOS at the 3 skeletal sites were significantly correlated (r = 0.28-0.44; P < 0.001). Significant correlations were found between SOS at all sites and BMD (r = 0.21-0.41; P < 0.001). The odds ratio for fracture prediction for SOS was 1.47 for tibia, 1.69 for radius, and 2.69 for phalanx. The corresponding odds ratios for BMD at the lumbar spine, femoral neck, and total hip ranged from 2.08 to 3.26. The area under the receiver operating characteristic curve ranged from 0.611 to 0.741 for SOS measurements and from 0.745 to 0.797 for BMD measurements.

CONCLUSIONS

The Omnisense multisite QUS device exhibits reproducible performance. Among the QUS variables, the phalangeal SOS provides the best discrimination of fracture patients.

Для цитирования: Damilakis John, Papadokostakis George, Vrahoriti Hrisanthi, Tsagaraki Ioanna, Perisinakis Kostas, Hadjipavlou Alexander, Gourtsoyiannis Nicholas. Ultrasound velocity through the cortex of phalanges, radius, and tibia in normal and osteoporotic postmenopausal women using a new multisite quantitative ultrasound device.. Investigative radiology. 2003. DOI: 10.1097/01.rli.0000057031.21810.f4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.