MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2196/50891

Медицинская библиотека · JMIR human factors · 2023

The Feasibility and Reliability of Upper Arm-Worn Apple Watch Heart Rate Monitoring for Surgeons During Surgery: Observational Study.

Yamada Kazunosuke, Enokida Yasuaki, Kato Ryuji, Imaizumi Jun, Takada Takahiro, Ojima Hitoshi

Наблюдательное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Health care professionals, particularly those in surgical settings, face high stress levels, impacting their well-being. Traditional monitoring methods, like using Holter electrocardiogram monitors, are impractical in the operating room, limiting the assessment of physicians' health. Wrist-worn heart rate monitors, like the Apple Watch, offer promise but are restricted in surgeries due to sterility issues.

OBJECTIVE

This study aims to assess the feasibility and accuracy of using an upper arm-worn Apple Watch for heart rate monitoring during robotic-assisted surgeries, comparing its performance with that of a wrist-worn device to establish a reliable alternative monitoring site.

METHODS

This study used 2 identical Apple Watch Series 8 devices to monitor the heart rate of surgeons during robotic-assisted surgery. Heart rate data were collected from the wrist-worn and the upper arm-worn devices. Statistical analyses included calculating the mean difference and SD of difference between the 2 devices, constructing Bland-Altman plots, assessing accuracy based on mean absolute error and mean absolute percentage error, and calculating the intraclass correlation coefficient.

RESULTS

The mean absolute errors for the whole group and for participants A, B, C, and D were 3.63, 3.58, 2.70, 3.93, and 4.28, respectively, and the mean absolute percentage errors were 3.58%, 3.34%, 2.42%, 4.58%, and 4.00%, respectively. Bland-Altman plots and scatter plots showed no systematic error when comparing the heart rate measurements obtained from the upper arm-worn and the wrist-worn Apple Watches. The intraclass correlation coefficients for participants A, B, C, and D were 0.559, 0.651, 0.508, and 0.563, respectively, with a significance level of P<.001, indicating moderate reliability.

CONCLUSIONS

The findings of this study suggest that the upper arm is a viable alternative site for monitoring heart rate during surgery using an Apple Watch. The agreement and reliability between the measurements obtained from the upper arm-worn and the wrist-worn devices were good, with no systematic error and a high level of accuracy. These findings have important implications for improving data collection and management of the physical and mental demands of operating room staff during surgery, where wearing a watch on the wrist may not be feasible.

Для цитирования: Yamada Kazunosuke, Enokida Yasuaki, Kato Ryuji, Imaizumi Jun, Takada Takahiro, Ojima Hitoshi. The Feasibility and Reliability of Upper Arm-Worn Apple Watch Heart Rate Monitoring for Surgeons During Surgery: Observational Study.. JMIR human factors. 2023. DOI: 10.2196/50891.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.