Loosening of a femoral stem associated with the use of an extended-lip acetabular cup liner. A case report.

Bosco J A, Benjamin J B

Клинический случайСтатьяПеревод в очереди

Аннотация

A case of femoral component loosening secondary to impingement on an extended-lip acetabular cup liner is presented. This impingement led to the accelerated creation of particulate polyethylene wear debris. The particulate polyethylene induced an osteolytic response about the femoral component, which contributed to the failure of this component. It is recommended that intraoperative examination for impingement, especially in extension, be performed when using liners of this design.

Для цитирования: Bosco J A, Benjamin J B. Loosening of a femoral stem associated with the use of an extended-lip acetabular cup liner. A case report.. The Journal of arthroplasty. 1993. DOI: 10.1016/s0883-5403(06)80113-0.