Complications of Semiconstrained Distal Radioulnar Joint Arthroplasty.

Bellevue Kate D, Thayer Mary K, Pouliot Michael, Huang Jerry I, Hanel Douglas P

СтатьяПеревод в очереди

Аннотация

PURPOSE: The Aptis total distal radioulnar joint (DRUJ) prosthesis is a semiconstrained implant designed for treatment of DRUJ arthritis and instability. The purpose of this study was to analyze short-term complications of this device.

METHODS

We performed a retrospective chart review of patients undergoing semiconstrained DRUJ arthroplasty from 2007 to 2015 at a single institution. Records were analyzed for complications and the need for subsequent surgical procedures.

RESULTS

Two senior hand surgeons at one institution performed 52 semiconstrained DRUJ arthroplasties over 8 years. Nineteen complications necessitating operative management occurred in 15 patients (29%). A total of 26 procedures were undertaken to address these complications. Complications included 4 periprosthetic fractures, 3 infections, 2 instances of aseptic loosening, 2 implant component failures, 1 instance of screw loosening, 3 neuromas requiring neurectomy, 2 instances of finger stiffness necessitating extensor tenolysis, and 2 cases of heterotopic ossification at the DRUJ. Three of the 52 implants were revised (6%) and 2 were explanted (4%); 3 of these (6%) were caused by deep infection.

CONCLUSIONS

There is limited literature on outcomes of the semiconstrained DRUJ prosthesis. Prior studies reported low complication rates, with 0% to 5% revisions. In the current clinical series, 29% of patients required further surgery for complications, the most common reasons for which were periprosthetic fracture and infection.

TYPE OF STUDY/LEVEL OF EVIDENCE: Therapeutic IV.

Для цитирования: Bellevue Kate D, Thayer Mary K, Pouliot Michael, Huang Jerry I, Hanel Douglas P. Complications of Semiconstrained Distal Radioulnar Joint Arthroplasty.. The Journal of hand surgery. 2017. DOI: 10.1016/j.jhsa.2017.11.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.