Comparison of cone beam computed tomography and plane radiographs of radial fractures as a basis for radiographical measurements.

Bry Kristian, Kortesniemi Mika, Koivikko Mika, Kerttula Liisa

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

The purpose of this study was to determine whether radiological measurements of radial fracture position made in cone beam computed tomography (CBCT) projection images are comparable to those made on traditional radiographs and could potentially substitute them.

METHODS

Sixteen patients with fractures of the distal radius referred for radiographs were recruited for an additional CBCT scan which was performed immediately afterwards. Projection images and volumetric data were saved from the CBCT scans. Measurements of ulnar variance, radial inclination and volar tilt were made from all three sets of images.

RESULTS

Agreement of projection image based measurements with radiographs was nearly as good as as the agreement of cross sectional image measurements with radiographs. The average difference between the results for projection images and radiographs were -1.2 mm (SD 1.9 mm), for radial inclination 0.7° (SD 2.9°) and for volar tilt 1.9° (SD 5.6°).

CONCLUSION

Differences between radiological measurements between the modalities studied are small and projection images could be used for the assessment of distal radial fractures.

Для цитирования: Bry Kristian, Kortesniemi Mika, Koivikko Mika, Kerttula Liisa. Comparison of cone beam computed tomography and plane radiographs of radial fractures as a basis for radiographical measurements.. BMC medical imaging. 2023. DOI: 10.1186/s12880-023-01093-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.