PURPOSE: How much can the radiation dose be reduced for hand radiography by using digital luminescence radiography (DLR)?

METHODS AND MATERIALS: A hand phantom (3M) with a cyst, two fractures and an "amputation" was digitally exposed in an anterior-posterior orientation using DLR (ADC-70, Agfa). The tube current time product (mAs) was reduced gradually while keeping the voltage constant. The surface entrance dose was measured by a Dosimax sensor (Wellhöfer). Five investigators evaluated the images for characteristics and critical features, pathological findings, visual resolution and contrast.

RESULTS

The surface entrance dose at 50 kV/5 mAs was 31 microGy. The images could be evaluated very well down to an average dose of 36% (11 microGy); sufficient images were obtained down to an average dose of 26% (8 microGy). The resolution of the line pairs was the same or reduced by 1 level depending on the investigator. Contrast was assessed as being very good to sufficient.

CONCLUSION

For the sufficient exclusion of a fracture the dose can be reduced to at least 39%, for the sufficient assessment of bony union and possible inflammatory changes to at least 42%, to position check and foreign body search to at least 23%. By DLR the following question-referred mAs-product at 50 kV are stated: fracture exclusion 2 mAs, bony union, arthritis and osteomyelitis 2.2 mAs, position check and foreign body search 1.25 mAs.

Для цитирования: Heyne J P, Merbold H, Sehner J, Neumann R, Adler R, Freesmeyer M, Kaiser W A. [Reduction of radiation dosage by using digital luminescence radiography on a hand phantom].. RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin. 2000. DOI: 10.1055/s-2000-336.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.