Biomechanical study of the stiffness of the femoral locking compression plate of an external fixator for lower tibial fractures.

Su Huan, Zhong Siyang, Ma Tianyong, Wu Weidong, Lu Yihong, Wang Dewei

СтатьяПеревод в очереди

Аннотация

BACKGROUND

A locking compression plate (LCP) of the distal femur is used as an external fixator for lower tibial fractures. However, in clinical practice, the technique lacks a standardized approach and a strong biomechanical basis for its stability.

METHODS

In this paper, internal tibial LCP fixator (Group IT-44), external tibial LCP fixator (Group ET-44), external distal femoral LCP fixator (Group EF-44, group EF-33, group EF-22), and conventional external fixator (Group CEF-22) frames were used to fix unstable fracture models of the lower tibial segment, and anatomical studies were performed to standardize the operation as well as to assess the biomechanical stability and adjustability of the distal femoral LCP external fixator by biomechanical experiments.

RESULTS

It was found that the torsional and flexural stiffnesses of group EF-44 and group EF-33 were higher than those of group IT-44 and group ET-44 (p < 0.05); the flexural stiffness of group EF-22 was similar to that of group IT-44 (p > 0.05); and the compressive stiffness of all three EF groups was higher than that of group ET-44 (p < 0.05). In addition, the flexural and compressive stiffnesses of the three EF groups decreased with the decrease in the number of screws (p < 0.05), while the torsional stiffness of the three groups did not differ significantly between the two adjacent groups (p > 0.05). Group CEF-22 showed the highest stiffnesses, while group ET-44 had the lowest stiffnesses (P < 0.05).

CONCLUSIONS

The study shows that the distal femoral LCP has good biomechanical stability and adjustability and is superior to the tibial LCP as an external fixator for distal tibial fractures, as long as the technique is used in a standardized manner according to the anatomical studies in this article.

Для цитирования: Su Huan, Zhong Siyang, Ma Tianyong, Wu Weidong, Lu Yihong, Wang Dewei. Biomechanical study of the stiffness of the femoral locking compression plate of an external fixator for lower tibial fractures.. BMC musculoskeletal disorders. 2023. DOI: 10.1186/s12891-023-06150-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.