STUDY

DESIGN

Retrospective clinical study.

OBJECTIVE

To evaluate the clinical outcome of patients who had undergone occipitocervical (OC) fusion using pedicle screws and rods over a minimum 5-year follow-up.

SUMMARY OF BACKGROUND DATA: Few studies have evaluated occipitocervical (OC) fusion using pedicle screws and rods for long-term follow-up.

METHODS

Twenty-seven consecutive patients treated underwent posterior OC fusion using pedicle screws and rods over a minimum 5-year follow-up. The Modified McCormick scale to grade a patient's functional status and the Japanese Orthopaedic Association (JOA) scoring system were used to evaluate preoperative and postoperative neurological function. We assessed fusion by both direct and indirect evidence; bony trabeculae at the graft-recipient interface on lateral cervical radiographs and sagittal computed tomography reconstruction was considered direct evidence of union.

RESULTS

The mean follow-up period was 7.2 years (5-14 y). JOA scores were 8.1±3.8 before surgery and 11.7±3.7 at the final follow-up. The recovery rate calculated from the JOA scores was 42.0±30.0%. Functional status did improve at least 1 grade according to the modified McCormick scale in 18 patients (66.7%). There was no deterioration at the final follow-up.There were postoperative implant-related complications in 8 patients (29.6%): loosening of pedicle screws in 2, rod breakage in 2, plate breakage in 1, screw breakage in 1, pullout of pedicle screws in 1, and wiring induced myelopathy in 1 patient. The average duration between surgery and implant failure was 31.2 months (12-60 mo) except for 2.

CONCLUSIONS

Sufficient bone grafting, proper decortication of the bone bed, using thicker and high stiffness rods, and ultra-high molecular weight polyethylene tape as a fixation or reinforcement of implant may help prevent implant failure.

Для цитирования: Ando Kei, Imagama Shiro, Ito Zenya, Kobayashi Kazuyoshi, Yagi Hideki, Shinjo Ryuichi, Hida Tetsuro, Ito Kenyu, Ishikawa Yoshimoto, Ishiguro Naoki. Minimum 5-year Follow-up Results for Occipitocervical Fusion Using the Screw-Rod System in Craniocervical Instability.. Clinical spine surgery. 2017. DOI: 10.1097/bsd.0000000000000199.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.