Fixation Strength in Full and Limited Fixation of Osteoporotic Distal Radius Fractures.
Аннотация
BACKGROUND
The purpose of this investigation is to determine whether osteoporotic intra-articular distal radius fractures surgically treated by filling all 7 distal screws of a volar plate will have a higher load to failure than those treated by filling only 4 distal screws.
METHODS
Ten matched pairs of fresh frozen cadaveric forearms were randomized within each pair to be treated by using either all 7 of the distal holes of a volar plate or only 4 distal screws. The distal radius fixation was performed with unicortical screws going to but not through the dorsal cortex, and the most distal screws were placed within 4 mm of the joint surface. An AO C2 type fracture was then created. All specimens were tested cyclically, with an axial load of 60 N, at 3 Hz for 1000 cycles to simulate early postoperative motion. All specimens were subsequently tested to mechanical failure.
RESULTS
There were no failures in either group during cyclic testing. There was no difference detected between groups for mean stiffness, yield load, peak load, or load to clinical failure. In both groups, the yield load, peak load, and load to clinical failure were higher than the 60- to 100-N forces encountered during postoperative rehabilitation.
CONCLUSIONS
There was no difference detected between osteoporotic intra-articular distal radius fractures treated by utilizing all 7 of the distal screws of a volar plate compared with those treated with only 4 distal screws.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.