PURPOSE: To compare the biomechanical properties of 6 dorsal and volar fracture fixation plate designs in a cadaver model.

METHOD: Six different plating techniques were used on surgically simulated, unstable, extra-articular distal radius fractures in fresh-frozen cadavers. Specimens were tested to failure in axial compression with the Materials Testing System machine, and were analyzed with a motion analysis system. The 6 different fixation systems studied included an AO stainless steel Pi plate (group 1), an AO titanium Pi plate (group 2), a Forte plate (group 3), a dorsally placed Symmetry plate (group 4), a volarly placed Symmetry plate (group 5), and a volarly placed SCS/V plate (group 6).

RESULTS

All dorsal plates (groups 1, 2, 3, 4) failed in apex dorsal angulation and all volar plates (groups 5, 6) failed in apex volar angulation. No group developed an average angular deformity greater than 5 degrees with a load of 100 N, which compares with the physiologic loads expected with active wrist motion. Only the volarly placed SCS/V plated specimens (group 6) resisted deformation of 5 degrees or more at loads up to 250 N, which compares with the physiologic loads expected with active finger motion, and was significantly stronger and more rigid than the other 5 plate groups.

CONCLUSIONS

The SCS/V plate fixation system is the most rigid of the systems tested and may offer adequate stability for the treatment of the distal radius fracture in which the anterior and/or posterior metaphyseal cortex is comminuted severely.

Для цитирования: Osada Denju, Viegas Steven F, Shah Munir A, Morris Randal P, Patterson Rita M. Comparison of different distal radius dorsal and volar fracture fixation plates: a biomechanical study.. The Journal of hand surgery. 2003. DOI: 10.1053/jhsu.2003.50016.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.