MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/1558944716628485

Медицинская библиотека · Hand (New York, N.Y.) · 2016

Comparison of Dorsal Plate Fixation Versus Intramedullary Headless Screw Fixation of Unstable Metacarpal Shaft Fractures: A Biomechanical Study.

Melamed Eitan, Hinds Richard M, Gottschalk Michael B, Kennedy Oran D, Capo John T

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

Background

Recently, intramedullary headless screw (IMHS) has shown promise as an alternative to dorsal plate fixation of metacarpal fractures. The purpose of this study was to assess the biomechanical performance of IMHS versus plating. We hypothesized that IMHS fixation provides inferior stability to plating.

Methods

Metacarpal fracture model with 3-mm of volar gapping in forty-four human cadaveric metacarpals was created. The specimens were divided into 5 groups: Group 1, 1.5-mm non-locking plate; Group 2, 1.5-mm locking plate; Group 3, 2.0-mm non-locking plate; Group 4, 2.0-mm locking plate; and Group 5, 2.4-mm short cannulated IMHS. A 4-point bending model was used to assess load-to failure (LTF) and stiffness.

Results

Mean LTF was 364 ± 130 N for 1.5-mm non-locking plates, 218 ± 94 N for 1.5-mm locking plates, 421 ± 86 N for 2.0-mm non-locking plates, 351 ± 71 N for 2.0-mm locking plates, and 75 ± 20 N for IMHS. Mean stiffness was 91 ± 12 N/mm for 1.5-mm non-locking plates, 110 ± 77 N/mm for 1.5-mm locking plates, 94 ± 20 N/mm for 2.0-mm non-locking plates, 135 ± 16 N/mm for 2.0-mm locking plates, and 55 ± 15 N/mm for IMHS. IMHS demonstrated significantly lower LTF and stiffness than plates.

Conclusions

IMHS fixation of unstable metacarpal shaft fractures offers less stability compared to plating when loaded in bending. The LTF and stiffness of IMHS versus plating of metacarpal shaft fractures has not been previously quantified. Our results reveal that IMHS fixation is less favorable biomechanically and should be carefully chosen in regards to fracture stability.

Для цитирования: Melamed Eitan, Hinds Richard M, Gottschalk Michael B, Kennedy Oran D, Capo John T. Comparison of Dorsal Plate Fixation Versus Intramedullary Headless Screw Fixation of Unstable Metacarpal Shaft Fractures: A Biomechanical Study.. Hand (New York, N.Y.). 2016. DOI: 10.1177/1558944716628485.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.