Anatomical Features can Affect OCT Measures Used for Clinical Decisions and Clinical Trial Endpoints.

Hood Donald C, La Bruna Sol, Durbin Mary, Lee Chris, Hsiao Yi S, De Moraes Carlos G, Tsamis Emmanouil

СтатьяПоддержка NIH: внешняя программаПеревод в очереди

Аннотация

PURPOSE: To understand the association between anatomical parameters of healthy eyes and optical coherence tomography (OCT) circumpapillary retinal nerve fiber layer (cpRNFL) thickness measurements.

METHODS

OCT cpRNFL thickness was obtained from 396 healthy eyes in a commercial reference database (RDB). The temporal quadrant (TQ), superior quadrant (SQ), inferior quadrant (IQ), and global (G) cpRNFL thicknesses were analyzed. The commercial OCT devices code these values based on percentiles (red, <1%; yellow, ≥1% and <5%), after taking age and disc area into consideration. Four anatomical parameters were assessed: fovea-to-disc distance, an estimate of axial length, and the locations of the superior and the inferior peaks of the cpRNFL thickness curve. Pearson correlation values were obtained for the parameters and the thickness measures of each of the four cpRNFL regions, and t-tests were performed between the cpRNFL thicknesses coded as abnormal (red or yellow, <5%) versus normal (≥5%).

RESULTS

For each of the four anatomical parameters, the correlation with the thickness of one or more of the TQ, SQ, IQ, and G regions exceeded the correlation with age or disc area. All four parameters were significantly (P < 0.001) associated with the abnormal cpRNFL values. The significant parameters were not the same for the different regions; for example, a parameter could be negatively correlated for the TQ but positively correlated with the SQ or IQ.

CONCLUSIONS

In addition to age and disc area, which are used for inferences in normative databases, four anatomical parameters are associated with cpRNFL thickness.

TRANSLATIONAL RELEVANCE: Taking these additional anatomical parameters into consideration should aid diagnostic accuracy.

Для цитирования: Hood Donald C, La Bruna Sol, Durbin Mary, Lee Chris, Hsiao Yi S, De Moraes Carlos G, Tsamis Emmanouil. Anatomical Features can Affect OCT Measures Used for Clinical Decisions and Clinical Trial Endpoints.. Translational vision science & technology. 2024. DOI: 10.1167/tvst.13.4.27.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.