The Effect of Screw Design and Cortical Augmentation on Insertional Torque and Compression in Coracoid-Glenoid Fixation in a Sawbones Model.

Rabinowitz Justin, Lin Jackie J, Greenhouse Alyssa, Rao Meghana V, Provencher Matthew, Parada Stephen, Friedman Richard J, Eichinger Josef K

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: To compare screw insertional torque and coracoid-glenoid compression from 4 fixation techniques with different screw design parameters and cortical augmentation for the Latarjet procedure.

METHODS

Simulated Latarjet procedures were performed with 4 fixation techniques using laminated polyurethane blocks with dimensions similar to the coracoid-glenoid construct. The groups included DePuy Synthes Mitek 3.5-mm partially threaded screws with top hats, Arthrex 3.75-mm fully threaded screws with a 2-hole plate, Arthrex 3.75-mm fully threaded screws, and Smith & Nephew 4.0-mm partially threaded screws. Screws were inserted using a digital torque-measuring screwdriver to determine maximum insertional torque. Pressure-sensitive film was used to measure the maximum contact pressure and the effective pressure distribution (EPD) between the coracoid and glenoid; the EPD represents the percentage of the film's surface area that experienced pressure greater than 10 MPa. One-way analysis of variance and post hoc tests were used for statistical analysis.

RESULTS

Significant differences were found between the 4 fixation groups for each variable measured. The 2 cortically augmented systems produced significantly higher maximum insertional torque than the non-cortically augmented systems (P < .001 for both). The 3.75-mm screws with a 2-hole plate yielded significantly higher contact pressures than the 4.0-mm screws (P = .028). This group also had a high EPD, with a mean value more than double the values of the non-cortically augmented systems (P = .037 and P < .001).

CONCLUSIONS

Cortically augmented fixation methods showed higher maximum insertional torque, maximum contact pressure, and EPD between the surfaces of the coracoid and glenoid in this Sawbones model.

CLINICAL RELEVANCE: Various implants are available for the Latarjet procedure, but their biomechanical characteristics have not yet been fully elucidated. Graft fracture and nonunion represent 2 modes of failure that may be related to insertional torque and coracoid-glenoid compression. This study compared screw insertional torque and compression achieved using 4 fixation techniques with different screw design parameters and cortical augmentation in a Sawbones model.

Для цитирования: Rabinowitz Justin, Lin Jackie J, Greenhouse Alyssa, Rao Meghana V, Provencher Matthew, Parada Stephen, Friedman Richard J, Eichinger Josef K. The Effect of Screw Design and Cortical Augmentation on Insertional Torque and Compression in Coracoid-Glenoid Fixation in a Sawbones Model.. Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association. 2019. DOI: 10.1016/j.arthro.2019.10.011.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.