DOI 10.1186/s12938-023-01083-1
Comparison of femoral neck system and three cannulated cancellous screws in the treatment of vertical femoral neck fractures: clinical observation and finite element analysis.
Аннотация
OBJECTIVE
The purpose of this study was to compare the biomechanical and clinical results of two surgical methods for the treatment of vertical femoral neck fractures: Femoral neck system (FNS) and traditional three cannulated cancellous screws (CCS).
METHODS
First, we developed three different vertical femoral neck fracture models for the finite element analysis, with angles of 55°, 65°, and 75°, respectively. Two experimental groups were set up: the FNS group and the CCS group. Each fracture group was tested under axial loads of 2100 N to measure the femur's displacement, Von Mises stress (VMS), and its internal fixation components. Secondly, we retrospectively included the cases of vertical femoral neck fractures with FNS and CCS in our hospital from May 2019 to May 2021. In this study, we compared the duration of intraoperative fluoroscopy, operative time, hospital stay, fracture healing time, Hemoglobin loss, Harris score of hip joint function, and postoperative complications among patients undergoing hip joint replacement.
RESULTS
In terms of finite element analysis, FNS has better anti-displacement stability than CCS at 55°and 65°, while FNS is greater than CCS in Von Mises stress. Clinically, we followed up on 87 patients for an average of 12 months. FNS was superior to traditional CCS in fracture healing time, operation time, fluoroscopy duration, fracture healing time, and Harris hip function score.
CONCLUSION
FNS is superior to traditional CCS in biomechanical and clinical aspects of treating vertical femoral neck fractures. There is potential for FNS to become a new treatment option for vertical femoral neck fractures.