MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1186/s12938-023-01083-1

Медицинская библиотека · Biomedical engineering online · 2023

Comparison of femoral neck system and three cannulated cancellous screws in the treatment of vertical femoral neck fractures: clinical observation and finite element analysis.

Huang Shaolong, Zhang Yazhong, Zhang Xu, Zhou Chengqiang, Li Wenbo, Wang Yunqing, Wang Bin, Zhu Ziqiang

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

The purpose of this study was to compare the biomechanical and clinical results of two surgical methods for the treatment of vertical femoral neck fractures: Femoral neck system (FNS) and traditional three cannulated cancellous screws (CCS).

METHODS

First, we developed three different vertical femoral neck fracture models for the finite element analysis, with angles of 55°, 65°, and 75°, respectively. Two experimental groups were set up: the FNS group and the CCS group. Each fracture group was tested under axial loads of 2100 N to measure the femur's displacement, Von Mises stress (VMS), and its internal fixation components. Secondly, we retrospectively included the cases of vertical femoral neck fractures with FNS and CCS in our hospital from May 2019 to May 2021. In this study, we compared the duration of intraoperative fluoroscopy, operative time, hospital stay, fracture healing time, Hemoglobin loss, Harris score of hip joint function, and postoperative complications among patients undergoing hip joint replacement.

RESULTS

In terms of finite element analysis, FNS has better anti-displacement stability than CCS at 55°and 65°, while FNS is greater than CCS in Von Mises stress. Clinically, we followed up on 87 patients for an average of 12 months. FNS was superior to traditional CCS in fracture healing time, operation time, fluoroscopy duration, fracture healing time, and Harris hip function score.

CONCLUSION

FNS is superior to traditional CCS in biomechanical and clinical aspects of treating vertical femoral neck fractures. There is potential for FNS to become a new treatment option for vertical femoral neck fractures.

Для цитирования: Huang Shaolong, Zhang Yazhong, Zhang Xu, Zhou Chengqiang, Li Wenbo, Wang Yunqing, Wang Bin, Zhu Ziqiang. Comparison of femoral neck system and three cannulated cancellous screws in the treatment of vertical femoral neck fractures: clinical observation and finite element analysis.. Biomedical engineering online. 2023. DOI: 10.1186/s12938-023-01083-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.