MedElevator Войти в исследования
← Поиск публикаций

DOI 10.21037/atm-22-4529

Медицинская библиотека · Annals of translational medicine · 2022

A finite element analysis of the supportive effect of a new type of rotary support plate on lateral tibial plateau fractures.

Gao Shijie, Yao Quan Cheng, Geng Lindan, Lu Jian, Li Ming, An Kai, Ren Guowei, Canavese Federico, Kim Seok Jung, Gwam Chukwuweike, Wang Pengcheng, Ren Dong

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Tibial plateau fractures (TPFs) are a challenging type of fracture in orthopedic traumatology. We previously designed a plate (Patent Number: CN201520195596.5) for posterolateral TPF combined with posterior lateral collapse.. In this study, finite element analysis was used to compare the biomechanical characteristics of two internal fixation methods for posterolateral TPF. We investigated the support effect of the new steel plate on lateral TPFs combined with posterior TPFs.

METHODS

Two models of complex TPF were established. Model A was fixed with the new type of plate, and model B was fixed without the plate. Three axial loads of 500, 1,000, and 1,500 N were applied using FEA on the two fracture models (A and B) to analyze the data.

RESULTS

In model A, the maximum displacement at 500, 1,000, and 1,500 N was 0.085797, 0.17043, and 0.25465 mm, respectively; the maximum stress of the bone block was 11.285, 20.648, and 29.227 MPa, respectively; and the maximum strain of the bone block was 0.0012474, 0.007435, and 0.0035769 mm, respectively. The maximum displacement of the internal fixation was 0.096932, 0.18682, and 0.27655 mm, respectively; the maximum stress was 69.54, 112.1, and 155.71 MPa, respectively; and the maximum strain was 0.00066228, 0.0010676, and 0.0014829 mm, respectively. In model B, the maximum displacement of fractures at 500, 1,000, and 1,500 N was 0.15675, 0.29868, and 0.44017 mm, respectively; the maximum stress of the bone block was 6.5519, 12.575, and 18.842 MPa, respectively; and the maximum strain of the bone block was 0.0032554, 0.0074357, and 0.012146 mm, respectively. The maximum displacement of the screw was 0.14177, 0.27109, and 0.39849 mm, respectively; the maximum stress was 48.916, 92.251, and 135.27 MPa, respectively; and the maximum strain was 0.00046608, 0.00087893, and 0.0012887 mm, respectively.

CONCLUSIONS

The fixation method using this type of plates and screws can replace other methods using two plates to fix complex TPF.

Для цитирования: Gao Shijie, Yao Quan Cheng, Geng Lindan, Lu Jian, Li Ming, An Kai, Ren Guowei, Canavese Federico, Kim Seok Jung, Gwam Chukwuweike, Wang Pengcheng, Ren Dong. A finite element analysis of the supportive effect of a new type of rotary support plate on lateral tibial plateau fractures.. Annals of translational medicine. 2022. DOI: 10.21037/atm-22-4529.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.