DOI 10.3390/bioengineering12050444
Modification of a Two-Part Cancellous Locking Screw: A Pilot Study on Increasing Resistance to Axial Pullout Strength.
Аннотация
Background/
Objectives
The pullout failure of conventional locking screws (LSs, screws with a locking mechanism) may occur in patients with osteoporosis, particularly when inserted near joints or across periarticular fractures (e.g., proximal humerus). The two-part locking cancellous screw modification (TP-LCS, screws composed of two parts) in metaphyseal cancellous bone is hypothesized to increase bone purchase and holding power. This study aimed to test the hypothesized advantages of TP-LCS over LSs.
Methods
An MTS 370 series frame with an axial/torsional load cell was used to test driving torque and axial pullout strength, following ASTM F543-07 standards. The TP-LCS group featured a newly modified screw design made from titanium alloy (Ti6Al4V), while conventional LSs (Synthes) were used for the control group. Statistical significance was assessed for selected comparisons relevant to the research objectives, including driving torque and axial pullout strength.
Results
The driving torque test showed that TP-LCS had a significantly higher maximum insertion torque (4.9 ± 0.4 N·cm) compared to LSs (4.2 ± 0.4 N·cm) (p = 0.0269), although no significant difference was found in maximum removal torque (p = 0.1046). The axial pullout test revealed that TP-LCS had significantly higher pullout strength (223.5 ± 12.2 N) compared to LSs (203.5 ± 11.5 N) (p = 0.0284). Failure during the axial pullout test often involved cracking of the test block material around the screw threads, causing the screw to pull out.
Conclusions
These results support the hypothesis that TP-LCS may offer improved axial pullout resistance compared to LSs, making it a potentially beneficial modification for LSs in osteoporotic metaphyseal regions or near joints. This study provides biomechanical insights into the advantages of the modified screw design over conventional LSs.