A biomechanical comparison of superior ramus plating versus intramedullary screw fixation for unstable lateral compression pelvic ring injuries,,

Hempen Eric C, Wheatley Benjamin M, Schimoler Patrick J, Kharlamov Alexander, Melvin Patricia R, Miller Mark Carl, Altman Gregory T, Altman Daniel T, Westrick Edward R

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

Management of the anterior component of unstable lateral compression (LC) pelvic ring injuries remains controversial. Common internal fixation options include plating and superior pubic ramus screws. These constructs have been evaluated in anterior-posterior compression (APC) fracture patterns, but no study has compared the two for unstable LC patterns, which is the purpose of this study.

METHODS

A rotationally unstable LC pelvic ring injury was modeled in 10 fresh frozen cadaver specimens by creating a complete sacral fracture, disruption of posterior ligaments, and ipsilateral superior and inferior rami osteotomies. All specimens were repaired posteriorly with two fully threaded 7 mm cannulated transiliac-transsacral screws through the S1 and S2 corridors. The superior ramus was repaired with either a 3.5 mm pelvic reconstruction plate (n = 5) or a bicortical 5.5 mm cannulated retrograde superior ramus screw (n = 5). Specimens were loaded axially in single leg support for 1000 cycles at 400 N followed by an additional 3 cycles at 800 N. Displacement and angulation of the superior and inferior rami osteotomies were measured with a three-dimensional (3D) motion tracker. The two fixation methods were then compared with Mann-Whitney U-Tests.

RESULTS

Retrograde superior ramus screw fixation had lower average displacement and angulation than plate fixation in all categories, with the motion at the inferior ramus at 800 N of loading showing a statistically significant difference in angulation.

CONCLUSION

Although management of the anterior ring in unstable LC injuries remains controversial, indications for fixation are becoming more defined over time. In this study, the 5.5 mm cannulated retrograde superior ramus screw significantly outperformed the 3.5 mm reconstruction plate in angulation of the inferior ramus fracture at 800 N. No other significance was found, however the ramus screw demonstrated lower average displacements and angulations in all categories for both the inferior and superior ramus fractures.

Для цитирования: Hempen Eric C, Wheatley Benjamin M, Schimoler Patrick J, Kharlamov Alexander, Melvin Patricia R, Miller Mark Carl, Altman Gregory T, Altman Daniel T, Westrick Edward R. A biomechanical comparison of superior ramus plating versus intramedullary screw fixation for unstable lateral compression pelvic ring injuries,,. Injury. 2022. DOI: 10.1016/j.injury.2022.09.027.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.