MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1136/bmjopen-2012-001541

Медицинская библиотека · BMJ open · 2013

The clinical implications of elevated blood metal ion concentrations in asymptomatic patients with MoM hip resurfacings: a cohort study.

Langton David J, Sidaginamale Raghavendra P, Joyce Thomas J, Natu Shonali, Blain Peter, Jefferson Robert Drysdale, Rushton Stephen, Nargol Antoni V F

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To determine whether elevated blood cobalt (Co) concentrations are associated with early failure of metal-on-metal (MoM) hip resurfacings secondary to adverse reaction to metal debris (ARMD).

DESIGN

Cohort study.

SETTING

Single centre orthopaedic unit.

PARTICIPANTS

Following the identification of complications potentially related to metal wear debris, a blood metal ion screening programme was instigated at our unit in 2007 for all patients with Articular Surface Replacement (ASR) and Birmingham MoM hip resurfacings. Patients were followed annually unless symptoms presented earlier. Symptomatic patients were investigated with ultrasound scan and joint aspiration. The clinical course of all 278 patients with 'no pain' or 'slight/occasional' pain and a Harris Hip Score greater than or equal to 95 at the time of venesection were documented. A retrospective analysis was subsequently conducted using mixed effect modelling to investigate the temporal pattern of blood Co levels in the patients and survival analysis to investigate the potential role of case demographics and blood Co levels as risk factors for subsequent failure secondary to ARMD.

RESULTS

Blood Co concentration was a positive and significant risk factor (z=8.44, p=2×10(-16)) for joint failure, as was the device, where the Birmingham Hip Resurfacing posed a significantly reduced risk for revision by 89% (z=-3.445, p=0.00005 (95% CI on risk 62 to 97)). Analysis using Cox-proportional hazards models indicated that men had a 66% lower risk of joint failure than women (z=-2.29419, p=0.0218, (95% CI on risk reduction 23 to 89)).

CONCLUSIONS

The results suggest that elevated blood metal ion concentrations are associated with early failure of MoM devices secondary to adverse reactions to metal debris. Co concentrations greater than 20 µg/l are frequently associated with metal staining of tissues and the development of osteolysis. Development of soft tissue damage appears to be more complex with females and patients with ASR devices seemingly more at risk when exposed to equivalent doses of metal debris.

Для цитирования: Langton David J, Sidaginamale Raghavendra P, Joyce Thomas J, Natu Shonali, Blain Peter, Jefferson Robert Drysdale, Rushton Stephen, Nargol Antoni V F. The clinical implications of elevated blood metal ion concentrations in asymptomatic patients with MoM hip resurfacings: a cohort study.. BMJ open. 2013. DOI: 10.1136/bmjopen-2012-001541.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.