MedElevator Войти в исследования
← Поиск публикаций

DOI 10.52312/jdrs.2022.676

Медицинская библиотека · Joint diseases and related surgery · 2022

Comparative finite element analysis of four different internal fixation implants for Pauwels type III femoral neck fractures in various fracture angles in the sagittal plane.

Özer Alaettin, Öner Kerim, Okutan Ahmet Emin, Ayas Muhammet Salih

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

This study aimed to evaluate the performance of four different fixation techniques for Pauwels type III femoral neck fractures considering the fracture morphology in the sagittal plane.

MATERIALS AND

METHODS

We constructed three different fracture morphologies in the sagittal plane in Pauwels type III femoral neck fractures: posteriorly angled at 20°, neutral, and anteriorly angled at 20°. We set up four fixation devices, including three cannulated screws (3CS), a dynamic hip screw with an antirotational screw (DHS+CS), a proximal femoral locking plate (PFLP), and three cannulated screws with a medial buttress plate (3CS+MBP). The twelve models were created and analyzed using the finite element analysis.

RESULTS

The finite element analysis revealed that 3CS+MBP yields better results in total vertical and rotational displacements, regardless of the fracture angle in the sagittal plane. For the anterior and posterior angled fractures in the sagittal plane, the PFLP was superior to the DHS+CS. However, the DHS+CS exhibited less displacement than the PFLP in the neutral fracture line in the sagittal plane. The 3CS group demonstrated poor mechanical stability for Pauwels type III fractures.

CONCLUSION

Regardless of the fracture line in the sagittal plane, the 3CS+MBP showed better biomechanical behaviors than the 3CS, DHS+CS, and PFLP. In addition, in contrast to the DHS+CS, the PFLP displayed less vertical and rotational displacement in the anterior and posterior fracture lines in the sagittal plane.

Для цитирования: Özer Alaettin, Öner Kerim, Okutan Ahmet Emin, Ayas Muhammet Salih. Comparative finite element analysis of four different internal fixation implants for Pauwels type III femoral neck fractures in various fracture angles in the sagittal plane.. Joint diseases and related surgery. 2022. DOI: 10.52312/jdrs.2022.676.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.