MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.cmpb.2022.107006

Медицинская библиотека · Computer methods and programs in biomedicine · 2022

Plate configuration for biological reconstructions of femoral intercalary defect - a finite element evaluation.

He Ziyang, Huang Siyi, Ji Tao, Tang Xiaodong, Yang Rongli, Guo Wei

СтатьяПеревод в очереди

Аннотация

BACKGROUND AND

OBJECTIVE

Biological reconstruction was commonly used for femoral intercalary defect. The initial stability by plate fixation was believed to have an effect on bone union and implant failure. Our study was proposed to explore relationship of plate configuration and initial stability for femoral intercalary reconstruction using allo-/autograft.

METHODS

Femoral intercalary defect models were established with four different plate configurations: (1) Single lateral bridging plate, SLP (2) Lateral bridging plate + Orthogonal adjuvant plate, LP+OAP (3) Lateral bridging plate + Medial adjuvant plate, LP+MAP (4) Lateral bridging plate + Medial bridging plate, LP+MP. A diaphysis defect of 12 cm was simulated, and the removed native femoral bone was used as a structural allograft with the osteotomy gap of 2 mm. Models were analyzed by finite element simulations under an axial compression of 2000N and an axial moment of 10 Nm, respectively.

RESULTS

Axial load: (1) The peak von Mises stress of SLP, LP+OAP, LP+MAP, LP+MP were 993.50 MPa, 335.63 MPa, 240.03 MPa, 281.73 MPa, respectively and LP+MAP was the lowest (p < 0.01); (2) The mean displacement of SLP, LP+OAP, LP+MAP, LP+MP was 0.765, 0.130, 0.121, 0.235 mm, respectively. LP+MAP showed the best stability while SLP had a crash in the medial proximal gap; (3) The LP+MAP configuration had the most uniform stress distribution and the lowest maximum von Mises stress of 79.7 MPa within plates. Axial torsional load: (1) The peak von Mises stress of SLP, LP+OAP, LP+MAP, LP+MP were 431.66Mpa, 120.73 MPa, 72.31 MPa, 109.86 MPa, respectively; (2) The rotation angle of SLP, LP+OAP, LP+MAP, LP+MP was 4.30°, 1.35°, 1.20°, 1.57°, respectively. All of LP+OAP, LP+MAP and LP+MP showed an optimal torsional stability.

CONCLUSIONS

For femoral intercalary reconstruction using allo-/autograft fixed by plates, LP+MAP and LP+MP configurations showed superior stability in terms of axial compression and torsion load by FE simulation. A better stability was believed to be associated with higher union rate and lower hardware failure rate.

Для цитирования: He Ziyang, Huang Siyi, Ji Tao, Tang Xiaodong, Yang Rongli, Guo Wei. Plate configuration for biological reconstructions of femoral intercalary defect - a finite element evaluation.. Computer methods and programs in biomedicine. 2022. DOI: 10.1016/j.cmpb.2022.107006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.