MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s11999-012-2430-4

Медицинская библиотека · Clinical orthopaedics and related research · 2012

Coordinating retrieval and register studies improves postmarket surveillance.

Ellison Peter, Hallan Geir, Høl Paul Johan, Gjerdet Nils Roar, Havelin Leif I

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The relative risk of revision of the Titan(®) femoral stem due to aseptic loosening increased after 2000; however, the reasons for this have not been established. A retrieval analysis was initiated with the aim of delineating the failure mechanism.

QUESTIONS/PURPOSES: We asked whether aseptic loosening in stems after 2000 was associated with (1) appearance of osteolytic lesions, (2) wear particle exposure, (3) stem damage, or (4) changes to the implant or surgical instrumentation.

METHODS

Femoral stems, cement, tissue, and radiographs were collected from 28 patients. We assessed the development of osteolytic lesions in 17 patients. Exposure to wear particles was quantified in 18 patients. Stem damage was assessed in 15 patients. We observed differences in the implants by examination of 24 retrieved stems. Information concerning changes to instrumentation was requested from the manufacturer.

RESULTS

We found osteolysis in all patients receiving implants after 2000, which was associated with a median dose of cement and stem particles of 14,726/mm(2). Abrasion covered 59% of the surface of stems implanted from 1999. We identified geometric changes to the stem, the percent weight of aluminum in the stem's oxide layer decreased from 25% to 14% after 1997 and the rasp used to prepare the femoral cavity changed to a broach in 1999.

CONCLUSIONS

Stems implanted from 2000 failed through osteolysis induced by particles released from the cement and implant. Changes to implant geometry, surface oxide layer, and surgical tools occurred in the same time frame as the reduction in survivorship.

Для цитирования: Ellison Peter, Hallan Geir, Høl Paul Johan, Gjerdet Nils Roar, Havelin Leif I. Coordinating retrieval and register studies improves postmarket surveillance.. Clinical orthopaedics and related research. 2012. DOI: 10.1007/s11999-012-2430-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.