DOI 10.11607/jomi.9657

Медицинская библиотека · The International journal of oral & maxillofacial implants · 2022

Fracture Resistance of Custom Lithium Disilicate Implant Restorations with Two Fabrication Techniques and Two Designs.

Emam Marwa, Eldimeery Ayman, Salah Tarek

СтатьяПеревод в очереди

Аннотация

PURPOSE: To compare the fracture resistance of a press-on ceramic custom implant restoration with pressed and cemented restorations.

MATERIALS AND

METHODS

Thirty-two (32) lithium disilicate (IPS e.max Press) custom hybrid abutment restorations were fabricated. The restorations were divided into two groups (n = 16) according to the construction technique: the commercial control group (C) and the press-on group (P). For the control group, lithium disilicate restorations were pressed and cemented on titanium bases. For the press-on group, lithium disilicate pressable ceramic (IPS e.max Press) was pressed on the titanium bases with injection molding. Each group was further divided according to the restoration design, either screw- or cement-retained, into two subgroups of eight specimens each. Specimens of C group were divided into screw-retained (cemented hybrid abutment crown, CHAC) or cement-retained (cemented hybrid abutment, CHA). Specimens of the P group were also divided into screw-retained (pressed hybrid abutment crown, PHAC) and cement-retained (pressed hybrid abutment, PHA). The specimens were subjected to static loading until failure with a universal testing machine. Two-way analysis of variance (ANOVA) was used to assess the effect of different techniques and designs on the fracture resistance of the samples (P < .05), followed by one-way ANOVA and Tukey honest significant difference (HSD) test (α = .05).

RESULTS

C group showed higher mean fracture resistance (812.443 ± 129.14 N) than P group (596.71 ± 108.83 N), and the difference was statistically significant (P < .05). Regarding restoration design, HA groups showed higher mean fracture resistance (742.621 ± 153.82 N) than HAC (666.53 ± 163.07 N) groups with no statistically significant difference. CHA showed the highest mean fracture resistance (817.65 ± 161.76 N), while PHAC showed the lowest mean fracture resistance values (525.83 ± 47.29 N).

CONCLUSION

The commercial cemented lithium disilicate restorations showed higher fracture resistance than the press-on restorations, although both showed a maximum load capacity that was greater than physiologic incisal force in the anterior region, and both hybrid abutments and hybrid abutment crowns were equally efficient in withstanding occlusal loading forces.

Для цитирования: Emam Marwa, Eldimeery Ayman, Salah Tarek. Fracture Resistance of Custom Lithium Disilicate Implant Restorations with Two Fabrication Techniques and Two Designs.. The International journal of oral & maxillofacial implants. 2022. DOI: 10.11607/jomi.9657.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.