BACKGROUND

Although early results with reverse total shoulder arthroplasty (rTSA) have been promising, concern exists about the high reported rates of scapular notching and the potential for catastrophic failure of glenoid component fixation. Generation of polyethylene wear debris may also contribute to notching and osteolysis of the scapula. A testing model for polyethylene wear is currently unavailable for reverse shoulder prostheses. The goal of this study was to develop a testing protocol using a commercially available hip simulator. Component design may also influence the generation of polyethylene debris. It is hypothesized that increased polyethylene wear occurs in glenospheres with holes in the articulating surface.

MATERIALS AND

METHODS

Custom fixtures were fabricated to simulate both glenohumeral abduction and flexion on a 12-station hip wear simulator. Loading profiles for both abduction and flexion were alternated every 250,000 cycles for a total of 5 million cycles. Gravimetric analysis of humeral cups throughout the test was used to characterize wear. Lubricant fluid was collected throughout the test and digested for polyethylene particle analysis.

RESULTS

Comparisons of volumetric wear rates and total volume loss between glenospheres with and without holes and between flexion and abduction loading profiles showed similar results. Particle analysis displayed fibrillar particles with an equivalent circle diameter of 0.3 ± .1 μm and an aspect ratio of 2.5 ± 1.4.

CONCLUSIONS

This study represents the first wear simulation and particle characterization of reverse shoulder systems. No significant difference in wear was reported between glenospheres with and without holes.

Для цитирования: Vaupel Zachary M, Baker Kevin C, Kurdziel Michael D, Wiater J Michael. Wear simulation of reverse total shoulder arthroplasty systems: effect of glenosphere design.. Journal of shoulder and elbow surgery. 2012. DOI: 10.1016/j.jse.2011.10.024.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.