DOI 10.1007/s00586-021-06860-x
The smartphone as a tool to screen for scoliosis, applicable by everyone.
Аннотация
PURPOSE (MAIN PURPOSES AND RESEARCH QUESTION): The purpose of this study is to assess the accuracy and precision of the smartphone with application and casing (Scolioscreen) compared to the Scoliometer.
METHODS
The Axial Trunk Rotation (ATR) was measured in adolescent scoliosis patients visiting the outpatient clinic while performing the Adam Forward Bending Test. The Scolioscreen measurements were performed by the orthopedic surgeon and a parent. They were compared to the measurement with the Scoliometer by the orthopedic surgeon, the gold standard. The accuracy was determined with the Pearson's correlation coefficient, and precision was determined by assessing the intra- and inter-variability with the intra-class correlation coefficient (ICC).
RESULTS
Fifty patients with adolescent idiopathic scoliosis (44 girls) were included with a mean age of 14.1 years and a mean Cobb angle of 38.5°. The accuracy of both the parents and orthopedic surgeon was excellent with a Pearson correlation coefficient of 0.92 and 0.97, respectively. All the ICC's, both intra- and inter-observer, were over 0.92 demonstrating excellent precision.
CONCLUSION
This study confirms the accuracy and precision of the Scolioscreen when measuring the ATR on patients with AIS. Therefore, the Scoliometer can be replaced by the more easily available Scolioscreen which can be used by both physician and parents.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.