BACKGROUND

Tibial component aseptic loosening remains problematic in primary total knee arthroplasty (TKA). Influential factors include component design, metallurgy, and cement technique. Additionally, reports advocate for longer tibial stem fixation in high body mass index (BMI) patients. We have utilized a single stem length modular titanium baseplate in patients regardless of BMI, bone quality, or malalignment. We report the survivorship of this implant with focus on the impact of elevated BMI and postoperative malalignment.

METHODS

We retrospectively reviewed patients who underwent TKA with a single modular titanium baseplate with a cruciate-shaped keel between 2004 and 2018. In total, 2,949 TKAs with a minimum of 1-year follow-up were included. The mean follow-up was 7 years. The primary outcome was component failure stratified by BMI and postoperative malalignment. High viscosity cement was utilized in all cases. Chi-squared and t-tests were used to compare outcome variables across groups.

RESULTS

Eighty-five implants (2.8%) were revised with 46 (1.6%) for aseptic loosening. Failure was not associated with BMI, gender, American Society of Anesthesiologists class, or Charlson Comorbidity Index. There was no difference in failure rate by BMI (P = .26) or by malalignment (outside of 3° from neutral mechanical axis) (P = .67). Age was associated with failure as patients with failed TKAs were younger (61 vs 65, P < .01).

CONCLUSION

This design of a specific modular titanium base plate with a cruciate-shaped keel and grit blast surface demonstrated 99% survivorship regardless of patient BMI or malalignment over 7-year follow-up period. Consistent cement technique with high viscosity cement indicates that component design remains an important variable impacting survivorship in TKA.

Для цитирования: Stevenson Kimberly L, Blackburn Brenna E, Da Silva Adrik Z, Erickson Jill A, Anderson Lucas A, Pelt Christopher E, Gililland Jeremy M, Peters Christopher L. High Survivorship of a Modular Titanium Baseplate Independent of Body Mass Index and Malalignment.. The Journal of arthroplasty. 2022. DOI: 10.1016/j.arth.2022.02.006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.