DOI 10.1016/j.clinbiomech.2021.105346
How do spine instrumentation parameters influence the 3D correction of thoracic adolescent idiopathic scoliosis? A patient-specific biomechanical study.
Аннотация
BACKGROUND
Patient-specific models promises to support the surgical decision-making process, particularly in adolescent idiopathic scoliosis. The present computational biomechanical study investigates how specific instrumentation parameters impact 3D deformity correction in thoracic scoliosis.
METHODS
1080 instrumentation simulations of a representative patient were run. The independent instrumentation parameters were: screw pattern, upper and lower instrumented vertebrae, rod curvature and rod stiffness. ANOVA and correlation analyses analyzed how the instrumentation parameters influenced the 3D correction.
FINDINGS: Coronal plane correction was affected by the lower instrumented vertebra and rod stiffness (explaining 84% and 11%, respectively, of its overall variance). The sagittal profile was controlled by rod curvature and the upper vertebra (56% and 36%). The transverse plane vertebral rotation was influenced by lower, upper instrumented vertebra and screw pattern (35%, 32% and 19%). The Cobb angle correction was strongly correlated with the number of fused vertebrae, particularly when grouped by the upper instrumented vertebra (r = -0.91) and rod stiffness (r = -0.73). Thoracic kyphosis was strongly correlated with the number of fused vertebrae grouped by rod curvature (r = 0.84). Apical vertebral rotation was moderately correlated with the number of fused vertebrae grouped by upper/lower instrumented vertebra (r = 0.55/0.58), although variations were minimal.
INTERPRETATION: Instrumenting the last vertebra touching the central sacral vertical line improves 3D correction. A trade-off between a more cranial vs. caudal upper instrumented vertebra, respectively beneficial for coronal/sagittal vs. transverse plane correction, is required. High rod stiffness, differential rod contouring, and screw pattern were effective for coronal correction, thoracic kyphosis, and axial vertebral derotation, respectively.