DOI 10.1186/s13018-022-02963-3
Comparison of the fixation ability between lag screw and bone plate for oblique metacarpal shaft fracture.
Аннотация
BACKGROUND
For oblique metacarpal shaft fracture, if anatomical reduction is achieved through conservative cast immobilization rather than stable fixation, bone malrotation can easily occur, resulting in severe loss in hand prehensile function. However, whether bone plate fixation or only lag screw fixation is more preferable remains unclear. Few studies have evaluated whether screw fixation can provide biomechanical fixation strength similar to bone plate fixation.
OBJECTIVE
We assessed the difference in fixation strength between fixtation with two lag screws and bone plate for oblique metacarpal shaft fractures.
MATERIALS AND
METHODS
We created oblique metacarpal shaft fractures on 21 artificial bones and fixated them using (1) double lag screw (2LS group), (2) regular plate (RP group), or (3) locked plate (LP group). To obtain the force-displacement data, a cantilever bending test was conducted for each specimen through a material testing machine. One-way analysis of variance and a Tukey test were conducted to compare the maximum fracture force and stiffness of the three fixation methods.
RESULTS
The maximum fracture force of the 2LS group (mean + SD: 153.6 ± 26.5 N) was significantly lower than that of the RP (211.6 ± 18.5 N) and LP (227.5 ± 10.0 N) groups (p < 0.001). However, no significant differences were discovered between the RP and LP groups. The coefficient of variation for the maximum fracture force of the 2LS group (17.3%) was more than twice as high as that of the RP (8.7%) and LP (4.4%) groups. In addition, the stiffness of the three fixation methods was similar.
CONCLUSION
Compared with bone plate fixation, double lag screw fixation yielded slightly lower maximum bearable fracture force but similar stiffness. Therefore, this technique could be used for treating oblique metacarpal shaft fractures. However, using double lag screw fixation alone is technically demanding and requires considerable surgical experiences to produce consistent results.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.