BACKGROUND

We previously reported the outcomes of a case-control study, at a minimum of five years of follow-up, comparing metal-on-metal and ceramic-on-polyethylene bearings for cementless primary hip arthroplasty in active patients below the age of fifty years. This report is an update on these groups after a minimum duration of follow-up of twelve years.

METHODS

Thirty-nine metal-on-metal cementless hip replacements with a 28-mm-diameter Metasul articulation were compared with a control group that included thirty-nine cementless ceramic-on-polyethylene hip replacements performed with a 28-mm-diameter head. The Metasul group included thirty patients with a mean age of forty years (range, twenty-three to forty-nine years), and the control group included thirty-two patients with a mean age of forty-one years (range, fifteen to forty-nine years). The groups were matched for age, activity level, preoperative Harris hip score, acetabular cup diameter, and indication for hip arthroplasty. All patients had a high level of activity, with 82% rated as grade IV or V according to the Devane scale.

RESULTS

After a mean duration of follow-up of thirteen years (twelve to fourteen years), only one hip (3%) had asymptomatic acetabular osteolysis and no hip (0%) had been revised in the metal-on-metal group, whereas eighteen hips (46%) had osteolysis and eleven hips (28%) had been revised because of wear or osteolysis in the ceramic-on-polyethylene group (p < 0.003). In the metal-on-metal group, the median Co concentration in the whole blood was 0.95 μg/L (0.4 to 4.8 μg/L) and the median Cr concentration was 1.2 μg/L (0.1 to 5.6 μg/L). The twelve-year survival rate (with reoperation for any reason as the end point) was 100% in the metal-on-metal group and 70% (95% confidence interval, 63% to 77%) in the ceramic-on-polyethylene group (p = 0.003).

CONCLUSIONS

After twelve to fourteen years of follow-up, metal-on-metal implants demonstrated better radiographic and survival results than ceramic-on-polyethylene implants in young, very active patients. Current wrought metal-on-metal implants with a 28-mm-diameter head and high carbide concentration did not produce the high rates of osteolysis and allergic reactions that may be observed with cast low-carbide metal-on-metal bearings after a shorter duration of follow-up.

Для цитирования: Migaud Henri, Putman Sophie, Krantz Nicolas, Vasseur Laurent, Girard Julien. Cementless metal-on-metal versus ceramic-on-polyethylene hip arthroplasty in patients less than fifty years of age: a comparative study with twelve to fourteen-year follow-up.. The Journal of bone and joint surgery. American volume. 2011. DOI: 10.2106/jbjs.j.01720.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.