Biomechanical comparison of acetabular fracture fixation with stand-alone THA or in combination with plating.

Wenzel Lisa, Sandriesser Sabrina, Glowalla Claudio, Gueorguiev Boyko, Perl Mario, Stuby Fabian M, Augat Peter, Hungerer Sven

СтатьяПеревод в очереди

Аннотация

PURPOSE: A common surgical treatment in anterior column acetabular fractures with preexisting osteoarthritis is THA, which is commonly combined with plate osteosynthesis. Implantation of a solitary revision cup cranially fixed to the os ilium is less common. The purpose of this study was to compare the stabilization of anterior column acetabular fractures fixed with a cranial socket revision cup with flange and iliac peg or with a suprapectineal plate osteosynthesis combined with an additional revision cup.

METHODS

In 20 human hemipelves, an anterior column fracture was stabilized by either a cranial socket revision cup with integrated flange (CF = Cup with Flange) or by a suprapectineal plate combined with a revision cup (CP = Cup and Plate). Each specimen was loaded under a stepwise increasing dynamic load protocol. Initial construct stiffness, interfragmentary movements along the fracture line, as well as femoral head movement in relation to the acetabulum were analyzed.

RESULTS

Both groups showed comparable initial construct stiffness (CP: 3180 ± 1162 N/mm and CF: 3754 ± 668 N/mm; p = 0.158). At an applied load of 1400 N, interfragmentary movements at the acetabular (p = 0.139) and the supraacetabular region (p = 0.051) revealed comparable displacement for both groups and remained below 1 mm. Femoral head movement in relation to the acetabulum also remained below 1 mm for both test groups (p = 0.260).

CONCLUSION

From a biomechanical point of view, both surgical approaches showed comparable fracture reduction in terms of initial construct stiffness and interfragmentary movement. The potential benefit of the less-invasive cranial socket revision cup has to be further investigated in clinical studies.

Для цитирования: Wenzel Lisa, Sandriesser Sabrina, Glowalla Claudio, Gueorguiev Boyko, Perl Mario, Stuby Fabian M, Augat Peter, Hungerer Sven. Biomechanical comparison of acetabular fracture fixation with stand-alone THA or in combination with plating.. European journal of trauma and emergency surgery : official publication of the European Trauma Society. 2022. DOI: 10.1007/s00068-021-01872-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.