Failure of dual plating in a complex open distal femur fracture as the result of major trauma.

Ward Thomas Robert William, Garala Kanai, Riemer Bryan

Клинический случайСтатьяПеревод в очереди

Аннотация

A 61-year-old polytrauma patient was admitted with a right distal comminuted metaphyseal femoral fracture with intra-articular extension (Orthopaedic Trauma Association 33C2.3 classification) among other injuries. Due to the high degree of comminution and massive bone loss, this was initially managed with a dual plating open reduction internal fixation. Dual plating has shown to be a superior fixation method than single variable angle locking compression plate (VA-LCP) plating providing greater fixation in metaphyseal bone. Our case reports the failure of dual plating which required removal of metalwork and subsequent fixation using intramedullary nail and plate technique. Failure of dual plating is not well documented in the literature. The most recent radiographs taken 15 months postrevision surgery show that the bone has started to heal with evidence of callus formation.

Для цитирования: Ward Thomas Robert William, Garala Kanai, Riemer Bryan. Failure of dual plating in a complex open distal femur fracture as the result of major trauma.. BMJ case reports. 2021. DOI: 10.1136/bcr-2021-245362.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.