DOI 10.3109/17453674.2010.519163

Медицинская библиотека · Acta orthopaedica · 2010

Measurement of femoral head penetration in polyethylene using a 3-dimensional CT-scan technique.

Vandenbussche Eric, Saffarini Mohammed, Hansen Ulrich, Taillieu Fabienne, Mutschler Céline, Augereau Bernard, Gregory Thomas M

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Current techniques for measuring in vivo polyethylene wear suffer from a range of problems, resulting in an unacceptable lack of repeatability and/or insufficient accuracy when they are used to measure the low wear rates associated with new, highly crosslinked polyethylene. We describe an improved CT method for measurement of 3D femoral head penetration in PE acetabular cups that has sufficient accuracy and repeatability to allow assessment of the wear potential of modern implants.

METHOD: The accuracy and repeatability of the CT-scan method was determined by blindly repeating measurements on a precisely calibrated 28-mm prosthetic head and by comparing them with direct metrological measurements on 10 acetabular specimens with in vitro wear from machining, and on 8 explanted acetabular specimens with in vivo wear.

RESULTS

The intra- and interobserver errors in femoral head diameter were 0.036 mm (SD 0.044) and 0.050 mm (SD 0.022), respectively. CT estimated femoral head penetration in both all-poly and metal-backed acetabular components with accuracy ranging from 0.009 to 0.245 mm (mean 0.080; SD 0.067).

INTERPRETATION: We found that the CT method is rapid, is accurate, and has repeatability and ease of availability. Using a slice thickness of 0.0625 mm, this method can detect wear—and also the threshold for the wear rate that causes osteolysis—much earlier than previous methods.

Для цитирования: Vandenbussche Eric, Saffarini Mohammed, Hansen Ulrich, Taillieu Fabienne, Mutschler Céline, Augereau Bernard, Gregory Thomas M. Measurement of femoral head penetration in polyethylene using a 3-dimensional CT-scan technique.. Acta orthopaedica. 2010. DOI: 10.3109/17453674.2010.519163.

Предметные рубрики MeSH