MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s10195-010-0104-0

Медицинская библиотека · Journal of orthopaedics and traumatology : official journal of the Italian Society of Orthopaedics and Traumatology · 2010

Direct comparison of polyethylene wear in cemented and uncemented acetabular cups.

Bjerkholt Håvard, Høvik Oystein, Reikerås Olav

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

It has been indicated that, in the long term, the rate of wear and the degree of osteolysis observed with uncemented acetabular components are greater than those associated with cemented cups, but most studies which compare the wear characteristics of cementless with cemented cups have used historical controls. We report a direct comparison of wear of a cemented and an uncemented cup with similar design, polyethylene, and sterilization method.

MATERIALS AND

METHODS

The study cohort includes 92 patients who were operated in 1997 with primary total hip replacement and have been followed for a period of 9-10 years. All patients were operated by posterolateral approach. In patients 70 years or older we used a cemented cup, in those 60 years or younger we used an uncemented cup, and in patients between 60 and 70 years we used either a cemented or uncemented cup as decided by the surgeon. At follow-up, radiographic imaging was obtained as standard anterioposterior view of the pelvis, and mean wear was determined as described by Livermore et al.

RESULTS

The overall wear of the cemented acetabular components was 1.07 ± 0.78 mm, and that of the uncemented cups was 1.18 ± 0.61 mm (P = 0.529). Wear was significantly associated with male sex (P = 0.003), younger age (P = 0.003), and degree of inclination (P < 0.001), but wear was not significantly associated with cemented versus uncemented cup (P = 0.437).

CONCLUSION

Our findings in this 9-10-year follow-up study suggest that cementless cups wear no more than cemented cups of similar design.

Для цитирования: Bjerkholt Håvard, Høvik Oystein, Reikerås Olav. Direct comparison of polyethylene wear in cemented and uncemented acetabular cups.. Journal of orthopaedics and traumatology : official journal of the Italian Society of Orthopaedics and Traumatology. 2010. DOI: 10.1007/s10195-010-0104-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.