DOI 10.1016/j.injury.2017.05.030
The learning curve with a new cephalomedullary femoral nail.
Аннотация
INTRODUCTION
The Cephalomedullary Nail (CMN) (Zimmer, Warsaw) was introduced in 2010 as part of a multicenter trial to evaluate its performance. At one year the CMN had results in keeping with other intramedullary devices with good union rates and low complication rates. In the second and third years of use an increased rate of implant failure was observed, towards the higher end of the 1-5% nail breakage rate seen in other studies. This study aims to evaluate if there any common features in this cohort of patients.
MATERIALS AND
METHODS
This is a retrospective cohort study looking at patients who underwent femoral fracture fixation using the cephalomedullary nail between January 2011 and June 2014. The primary outcome measure was implant failure; secondary outcomes were; fracture reduction and bisphosphonate use.
RESULTS
201 patients were included (135 female, 66 male) with an average age of 81 years. Ten (5%) nail breakages occurred in the study period at an average of 39 weeks (24-92); 9 were 125° nails 1 was a 130° nail and all fractured at the lag screw junction.
CONCLUSIONS
Implant failure is a recognised complication of intramedullary nailing in cases of non-union. The increased rate of implant failure in our department required a change to a 130° CMN implant and a 3.2mm diameter guide wire for placement of the lag screw. We continue to monitor this difficult group of patients very closely.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.