Second-generation highly cross-linked X3™ polyethylene wear: a preliminary radiostereometric analysis study.

Campbell David G, Field John R, Callary Stuart A

СтатьяПеревод в очереди

Аннотация

BACKGROUND

First-generation highly cross-linked polyethylene liners have reduced the incidence of wear particle-induced osteolysis. However, failed acetabular liners have shown evidence of surface cracking, mechanical failure, and oxidative damage. This has led to the development of second-generation highly cross-linked polyethylene, which has improved wear and mechanical properties and resistance to oxidation in vitro. Owing to its recent introduction, there are no publications describing its clinical performance.

QUESTIONS/PURPOSES: We assessed early clinical wear of a second-generation highly cross-linked polyethylene liner and compared its clinical performance with the published results of hip simulator tests and with first-generation highly cross-linked polyethylene annealed liners.

PATIENTS AND

METHODS

Twenty-one patients were enrolled in a prospective cohort study. Clinical outcome and femoral head penetration were measured for 19 patients at 6 months and 1 and 2 years postoperatively.

RESULTS

The median proximal head penetration was 0.009 mm and 0.024 mm at 1 and 2 years, respectively. The median two-dimensional (2-D) head penetration was 0.083 mm and 0.060 mm at 1 and 2 years, respectively. The median proximal wear rate between 1 and 2 years was 0.015 mm/year.

CONCLUSIONS

The wear rate calculated was similar to the in vitro wear rate reported for this material; however, it was less than the detection threshold for this technique. Although longer followup is required for wear to reach a clinically quantifiable level, this low level of wear is encouraging for the future clinical performance of this material.

LEVEL OF EVIDENCE: Level IV, therapeutic study. See the Guidelines for Authors for a complete description of levels of evidence.

Для цитирования: Campbell David G, Field John R, Callary Stuart A. Second-generation highly cross-linked X3™ polyethylene wear: a preliminary radiostereometric analysis study.. Clinical orthopaedics and related research. 2010. DOI: 10.1007/s11999-010-1259-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.